Catheter-induced pulmonary artery perforation. Mechanisms, management, and modifications

  • PMID: 7242132
Case Reports

Catheter-induced pulmonary artery perforation. Mechanisms, management, and modifications

P G Barash et al. J Thorac Cardiovasc Surg. 1981 Jul.

Abstract

Six cases of pulmonary artery perforation associated with the use of Swan-Ganz catheters are reviewed. Risk factors included pulmonary hypertension, anticoagulation, and hypothermia. The mechanisms leading to perforation were clarified by the use of postmortem studies employing isolated whole lung preparations. These studies revealed that perforation results from (1) tip perforation of vasculature, (2) eccentric balloon configuration propelling the balloon through the vessel wall, and (3) balloon inflation disrupting the pulmonary artery (mean intraballoon pressure 250 mm Hg). Early clinical symptoms include hemoptysis of bright red blood and/or hypotension. Immediate evaluation may necessitate examination with a fiberoptic bronchoscope and "wedge" angiogram. If massive hemoptysis occurs, isolation of the unaffected lung by endobronchial intubation is mandatory. Pneumonectomy or lobectomy may be required. Revised guidelines for catheter insertion and pulmonary capillary wedge pressure (PCWP) measurements are presented. Finally, consideration is given to redesigning the pulmonary artery flow-guided catheter, particularly for use in patients undergoing cardiac operations with systemic anticoagulation. Modifications should be directed at (1) softer catheter tip with temperature-insensitive body, (2) low-pressure balloon, and (3) balloon pressure relief valve.

PubMed Disclaimer

Publication types

LinkOut - more resources