A comparative study of automated blood pressure recorders
- PMID: 7267490
- PMCID: PMC2424836
- DOI: 10.1136/pgmj.56.662.815
A comparative study of automated blood pressure recorders
Abstract
Four automated blood pressure machines (Arteriosonde 1225, Bosomat 11D, Elag Koln BE 237R and Omron HEM 3) have been compared with each other and with the Hawksley Random Zero Sphygmomanometer in 18 subjects with a wide range of BP. Each form of measurement provided a high degree of internal consistency. The Arteriosonde 1225 consistently recorded systolic and diastolic pressures which were below the values obtained with the Random Zero instrument (P=0.001). The other 3 automated machines recorded BPs which did not differ significantly from the Random Zero or from each other. Heart rate readings did not differ significantly for any of the machines compared to manual recording. The disparity attributable to the Arterisonde could result from its being the only form of measurement to use an ultrasonic rather than a microphonic recording technique. However, the cheaper and more portable automated machines produce results which closely reflect those obtained by using a mercury sphygmomanometer.
Similar articles
-
Using Hawksley random zero sphygmomanometer as a gold standard may result in misleading conclusions.Blood Press. 1994 Sep;3(5):283-6. doi: 10.3109/08037059409102275. Blood Press. 1994. PMID: 7866591
-
Does the Hawksley random zero sphygmomanometer underestimate blood pressure, and by how much?J Hum Hypertens. 1995 May;9(5):337-43. J Hum Hypertens. 1995. PMID: 7623370 Clinical Trial.
-
[Are automatic devices suitable for blood pressure determinations?].Schweiz Med Wochenschr. 1979 Sep 1;109(33):1225-30. Schweiz Med Wochenschr. 1979. PMID: 493912 German.
-
Inaccuracy of the Hawksley random zero sphygmomanometer.Lancet. 1990 Dec 15;336(8729):1465-8. doi: 10.1016/0140-6736(90)93177-q. Lancet. 1990. PMID: 1979092 Clinical Trial.
-
The Dinamap 1846SX automated blood pressure recorder: comparison with the Hawksley random zero sphygmomanometer under field conditions.J Epidemiol Community Health. 1992 Apr;46(2):164-9. doi: 10.1136/jech.46.2.164. J Epidemiol Community Health. 1992. PMID: 1583434 Free PMC article. Clinical Trial.
Cited by
-
A proposal for scientific validation of instruments for indirect blood pressure measurement at rest, during exercise, and in critical care.J Clin Monit. 1994 May;10(3):163-77. doi: 10.1007/BF02908856. J Clin Monit. 1994. PMID: 8027747
-
Pharmacodynamic studies on mianserin and its interaction with clonidine.Eur J Clin Pharmacol. 1981;21(2):97-102. doi: 10.1007/BF00637508. Eur J Clin Pharmacol. 1981. PMID: 7341285 Clinical Trial.
-
Absence of an effect of mianserin on the actions of clonidine or methyldopa in hypertensive patients.Eur J Clin Pharmacol. 1983;24(1):15-9. doi: 10.1007/BF00613921. Eur J Clin Pharmacol. 1983. PMID: 6299752 Clinical Trial.