. 1981 Jul-Aug;16(4):337-41.
doi: 10.1097/00004424-198107000-00017.

Factors affecting the measurement of size and CT number in computed tomography

Factors affecting the measurement of size and CT number in computed tomography

B S Baxter et al. Invest Radiol. 1981 Jul-Aug.

Abstract

Computed tomography scanners may be used to make anatomic size measurement; however, manipulating the viewer controls (especially the window center) can have a significant influence on the apparent size of structures in the image. A special phantom was constructed to study the effect of window center adjustments on the apparent size of lone cylindrical objects (no variation within slice thickness) and also for spherical objects (maximum variation within slice thickness). Using this phantom, the authors were able to identify several commonly encountered situations in which both the apparent size and CT numbers may be seriously in error. It was found that for cylindrical objects larger than about one transverse resolution element and aligned with the scanner axis, CT numbers may be determined accurately; however, the apparent diameter changes by several millimeters as the window center setting is changed. CT numbers for spheres are inaccurate when the diameter is comparable to or less than the slice thickness and, as a result, diameters of spheres cannot be measured accurately with fixed window center settings. Diameters of spheres can be measured accurately (+/- 1 mm) if the sphere is centered in the slice and the full width at half maximum of the CT number profile is used.

PubMed Disclaimer

LinkOut - more resources