Measurements of computed tomography x-ray fields utilizing the partial volume effect
- PMID: 7393165
- DOI: 10.1118/1.594696
Measurements of computed tomography x-ray fields utilizing the partial volume effect
Abstract
Due to the x-ray field geometry, exposure measurements on computed tomographic (CT) scanners require modification of widely used measurement techniques. Ionization chambers have geometries which do not allow measurement of narrow rotating x-ray fields. However, by utilizing the partial volume effect, an ion chamber can be calibrated to measure narrow x-ray fields with a spatial resolution of down to 1 mm. The number of scans required to reconstruct a complete profile depends on the resolution required, and is typically on the order of 10 to 20 scans. This is accomplished using a series of attenuating sleeves over the ionization chamber. The chamber and sleeves are calibrated against a standard chamber using large field geometries. Comparisons with TLD measurements are shown to verify the accuracy of the technique.
Similar articles
-
Ionization chamber volume determination and quality assurance using micro-CT imaging.Phys Med Biol. 2008 Sep 21;53(18):5029-43. doi: 10.1088/0031-9155/53/18/012. Epub 2008 Aug 22. Phys Med Biol. 2008. PMID: 18723930
-
Concepts for dose determination in flat-detector CT.Phys Med Biol. 2008 Jul 7;53(13):3551-66. doi: 10.1088/0031-9155/53/13/011. Epub 2008 Jun 13. Phys Med Biol. 2008. PMID: 18552416
-
The measurement of radiation dose profiles for electron-beam computed tomography using film dosimetry.Med Phys. 1994 Aug;21(8):1287-91. doi: 10.1118/1.597200. Med Phys. 1994. PMID: 7799873
-
Multisegmented ion chamber for CT scanner dosimetry.Med Phys. 1981 Sep-Oct;8(5):640-5. doi: 10.1118/1.595022. Med Phys. 1981. PMID: 7290016
-
Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry.Med Phys. 2015 Aug;42(8):4954-64. doi: 10.1118/1.4927059. Med Phys. 2015. PMID: 26233221
Cited by
-
Techniques and parameters for estimating radiation exposure and dose in cardiac computed tomography.Int J Cardiovasc Imaging. 2005 Feb;21(1):165-76. doi: 10.1007/s10554-004-5338-6. Int J Cardiovasc Imaging. 2005. PMID: 15915949 Review.
-
Optimization-based algorithm for solving the discrete x-ray transform with nonlinear partial volume effect.J Med Imaging (Bellingham). 2020 Sep;7(5):053502. doi: 10.1117/1.JMI.7.5.053502. Epub 2020 Oct 6. J Med Imaging (Bellingham). 2020. PMID: 33033733 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical