Continuous transcutaneous oxygen monitoring during respiratory failure, cardiac decompensation, cardiac arrest, and CPR. Transcutaneous oxygen monitoring during arrest and CPR
- PMID: 7408502
- DOI: 10.1097/00003246-198007000-00003
Continuous transcutaneous oxygen monitoring during respiratory failure, cardiac decompensation, cardiac arrest, and CPR. Transcutaneous oxygen monitoring during arrest and CPR
Abstract
The transcutaneous oxygen sensor (PtcO2), which has been used to predict PaO2 in neonates, recently has been shown to follow changes in oxygen delivery, rather than PaO2 during shock and hypoxia in dogs. Six preterminal patients were continuously monitored with PtcO2 and monitored hemodynamically at frequent intervals during cardiac decompensation, arrest, and cardiopulmonary resuscitation (CPR). The weighted mean correlation coefficients between PtcO2 and O2 delivery as well as between PtcO2 and cardiac output were 0.94 and 0.96, respectively. Five patients died of severe ARDS and 1 patient died intraoperatively of hemorrhagic shock. Four patients were monitored 1-7 days before shock occurred. The correlation between PtcO2 and PaO2 was 0.91 during periods of normal cardiac output in the preterminal period. During cardiac decompensation, the cardiac output, PtcO2, and mixed venous oxygen tension (PcO2) of 25 torr was reached, the PtcO2 fell below the PvO2. This also corresponded to a decrease in VO2. The mean VO2 was 142 +/- 24 ml/min x M2 for PtcO2 values > torr, and 75 +/- 15 ml/min x M2 for PtcO2 < 25 torr (p < 0.01). A PtcO2 of > 40 torr corresponded to normal cardiac index, O2 delivery, VO2, PvO2, and arterial pH (pHa) while a PtcO2, of < 25 torr corresponded to large reductions of these variables. A PtcO2 of < 25 torr preceded cardiac arrest by 43 +/- 28 min.
Similar articles
-
Transcutaneous oxygen monitoring of critically ill adults, with and without low flow shock.Crit Care Med. 1981 Oct;9(10):706-9. doi: 10.1097/00003246-198110000-00007. Crit Care Med. 1981. PMID: 7285607
-
Noninvasive monitoring in emergency resuscitation.Ann Emerg Med. 1986 Dec;15(12):1434-6. doi: 10.1016/s0196-0644(86)80937-4. Ann Emerg Med. 1986. PMID: 3777616
-
Conjunctival and transcutaneous oxygen monitoring during cardiac arrest and cardiopulmonary resuscitation.Crit Care Med. 1984 May;12(5):419-21. doi: 10.1097/00003246-198405000-00001. Crit Care Med. 1984. PMID: 6713911
-
Conjunctival oxygen monitoring during cardiopulmonary resuscitation.Arch Emerg Med. 1989 Jun;6(2):128-36. doi: 10.1136/emj.6.2.128. Arch Emerg Med. 1989. PMID: 2662992 Free PMC article. Review.
-
Transcutaneous PO2 measurement.Can Anaesth Soc J. 1984 Nov;31(6):664-77. doi: 10.1007/BF03008765. Can Anaesth Soc J. 1984. PMID: 6388743 Review.
Cited by
-
Transorally obtained oxygen tension as an indicator of arterial oxygen tension.J Clin Monit. 1988 Apr;4(2):86-90. doi: 10.1007/BF01641807. J Clin Monit. 1988. PMID: 3373255
-
Monitoring the tissue perfusion during hemorrhagic shock and resuscitation: tissue-to-arterial carbon dioxide partial pressure gradient in a pig model.J Transl Med. 2021 Nov 14;19(1):390. doi: 10.1186/s12967-021-03060-5. J Transl Med. 2021. PMID: 34774068 Free PMC article.
-
A stochastic control program to predict outcome and to support therapeutic decisions: a preliminary report.J Clin Monit Comput. 2005 Jun;19(3):223-30. doi: 10.1007/s10877-005-5870-5. J Clin Monit Comput. 2005. PMID: 16244846
-
Factors influencing transcutaneous oxygen and carbon dioxide measurements in adult intensive care patients.Intensive Care Med. 1991;17(5):272-5. doi: 10.1007/BF01713936. Intensive Care Med. 1991. PMID: 1939871
-
Transcutaneous oxygen monitoring of patients undergoing surgical removal of wisdom teeth.Anesth Prog. 1982 May-Jun;29(3):70-1. Anesth Prog. 1982. PMID: 6956244 Free PMC article. No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical