Neonatal mercury vapor exposure in an infant incubator
- PMID: 7454493
Neonatal mercury vapor exposure in an infant incubator
Abstract
A healthy newborn infant acquired a substantial body burden of mercury by inhaling mercury vapor originating from a broken mercury expansion switch in the heating unit of an infant incubator. Highly toxic mercury vapor is produced in quantity by heating otherwise harmless metallic mercury. Switches and thermometers that contain mercury should be removed from infant incubators.
Publication types
MeSH terms
Substances
LinkOut - more resources