Morphology of Fallopian tubes removed from a patient after failure of clip sterilization

Case Reports

Morphology of Fallopian tubes removed from a patient after failure of clip sterilization

K A Walz et al. Arch Gynecol. 1980.

Abstract

A bilateral salpingectomy was performed at the time of vaginal hysterectomy in a pregnant patient sterilized 20 months earlier by application of Bleier clips to the fallopian tubes. Clip application to the left tube had been incomplete. Undisturbed tissue (left tube) and clipped tissue (right tube) were examined and compared by light and scanning electron microscopy. The segment of tube fully grasped by the clip had a completely detached lining epithelium with coarse and flattened mucosa, a loose stroma, and disorganized muscular bundles. The left tubal mucosa was normal, whereas the right tube was stenosed at the site of clip placement, with rigid walls and a 0.175 mm luminal diameter. With progressively greater stenosis toward the clip site, mucosal destruction increased. Polypoidal mucosal folds were seen, as well as fibrous adhesions between mucosal folds. These observations indicate that tissue damage is extensive enough after clip application to require excision of the damaged segment of tube and microsurgical re-anastomosis for reversal of sterilization.

PIP: A 37-year-old woman had a bilateral salpingectomy performed at the time of vaginal hysterectomy although she had been sterilized 20 months before via application of Bleier clips to the fallopian tubes. The woman presented pregnant. Using both a scanning electron and a light microscope, undisturbed tissue (left tube where clip application had been incomplete) and clipped tissue were examined. The right tube (clipped) showed completely detached lining epithelium with coarse and flattened mucosa, loose stroms, and disorganized muscular bundles in the segment where the clip had been attached. The left tubal mucosa, where the clip had failed to attach, was perfectly normal, whereas the right tube was stenosed at the site of clip placement. The tissue destruction seen after clip sterilization is so extensive that excision of damaged segment followed by microsurgical reanastomosis seems the only procedure to reverse sterilization of this type.

PubMed Disclaimer

Similar articles

References

    1. Geburtshilfe Frauenheilkd. 1978 Apr;38(4):299-303 - PubMed
    1. J Microsc. 1971 Jun;93(3):247-50 - PubMed
    1. Fertil Steril. 1975 Nov;26(11):1122-31 - PubMed
    1. Geburtshilfe Frauenheilkd. 1979 Jul;39(7):545-52 - PubMed
    1. Am J Obstet Gynecol. 1979 Dec 15;135(8):1016-20 - PubMed

Publication types

LinkOut - more resources