The efficacy of intramedullary femoral alignment in total knee replacement
- PMID: 7497656
The efficacy of intramedullary femoral alignment in total knee replacement
Abstract
The angle formed between the femoral anatomic axis and the femoral prosthesis was measured on radiographs of 201 total knee arthroplasties. The radiographs were taken on a 51 inch x 14 inch cassette and had to have satisfactory centering and minimal rotation to be included. All total knee arthroplasties had intramedullary femoral alignment. The mean femoral component angle (relative to the femoral anatomic axis) was 6.1 degrees valgus, with a standard deviation of 2 degrees and a range from 2 degrees to 12 degrees valgus. Seventeen knees (8.5%) had a femoral component angle that was < or = 2 degrees valgus or > or = 10 degrees valgus. In 4 knees, medial femoral bowing of the distal third of the femoral shaft caused an excessive valgus cut. In 7 knees, a capacious femoral canal was present that potentially allowed for divergence of the relatively undersized guide rod in the canal. Overall, 8.5% of cuts were believed to be less than ideal. Preoperative radiographs of the entire femur should identify patients in whom errors are more likely and allow for templating the amount of distal femoral bone to be resected. Extra caution is needed in bowed femora and in capacious femoral canals.
Similar articles
-
Accuracy of intramedullary versus extramedullary tibial alignment cutting systems in total knee arthroplasty.Clin Orthop Relat Res. 1995 Dec;(321):106-10. Clin Orthop Relat Res. 1995. PMID: 7497654
-
The effect of surgeon experience on component positioning in 673 Press Fit Condylar posterior cruciate-sacrificing total knee arthroplasties.J Arthroplasty. 2001 Aug;16(5):635-40. doi: 10.1054/arth.2001.23569. J Arthroplasty. 2001. PMID: 11503124
-
Femoral shaft bowing in the coronal plane has more significant effect on the coronal alignment of TKA than proximal or distal variations of femoral shape.Knee Surg Sports Traumatol Arthrosc. 2015 Jul;23(7):1936-42. doi: 10.1007/s00167-014-3006-5. Epub 2014 Apr 24. Knee Surg Sports Traumatol Arthrosc. 2015. PMID: 24760162
-
What is the optimal valgus pre-set for intramedullary femoral alignment rods in total knee arthroplasty?Knee Surg Sports Traumatol Arthrosc. 2017 Nov;25(11):3480-3487. doi: 10.1007/s00167-016-4141-y. Epub 2016 May 6. Knee Surg Sports Traumatol Arthrosc. 2017. PMID: 27154280
-
Seven cuts to the perfect total knee.Orthopedics. 2009 Sep;32(9):orthosupersite.com/view.asp?rID=42848. doi: 10.3928/01477447-20090728-27. Orthopedics. 2009. PMID: 19751004 Review.
Cited by
-
[Post-traumatic osteoarthritis of the knee: special indication for navigated TKA].Unfallchirurg. 2008 Sep;111(9):754-8. doi: 10.1007/s00113-007-1391-7. Unfallchirurg. 2008. PMID: 18301874 German.
-
Accuracy of implant placement utilizing customized patient instrumentation in total knee arthroplasty.Adv Orthop. 2013;2013:891210. doi: 10.1155/2013/891210. Epub 2013 Sep 16. Adv Orthop. 2013. PMID: 24151556 Free PMC article.
-
Short knee radiographs can be inadequate for estimating TKA alignment in knees with bowing.Knee Surg Relat Res. 2020 Jan 1;32(1):9. doi: 10.1186/s43019-019-0020-4. Knee Surg Relat Res. 2020. PMID: 32660638 Free PMC article.
-
The immediate effect of navigation on implant accuracy in primary mini-invasive unicompartmental knee arthroplasty.Knee Surg Sports Traumatol Arthrosc. 2008 Dec;16(12):1133-40. doi: 10.1007/s00167-008-0618-7. Epub 2008 Sep 20. Knee Surg Sports Traumatol Arthrosc. 2008. PMID: 18810391
-
Computer-assisted surgery can reduce blood loss after total knee arthroplasty.Knee Surg Sports Traumatol Arthrosc. 2009 Apr;17(4):356-60. doi: 10.1007/s00167-008-0683-y. Epub 2008 Dec 16. Knee Surg Sports Traumatol Arthrosc. 2009. PMID: 19083205 Clinical Trial.
MeSH terms
LinkOut - more resources
Full Text Sources