Development and validation of a high-sensitivity assay for an antipsychotic agent, CP-88,059, with solid-phase extraction and narrow-bore high-performance liquid chromatography
- PMID: 7550970
- DOI: 10.1016/0378-4347(95)00071-p
Development and validation of a high-sensitivity assay for an antipsychotic agent, CP-88,059, with solid-phase extraction and narrow-bore high-performance liquid chromatography
Abstract
An analytical method has been developed and validated for the quantitation of CP-88,059 in human serum. The compound and internal standard were extracted from serum by solid-phase extraction with a weak cation-exchange phase. The analytes were resolved from endogenous interferences using narrow-bore (2.1 mm I.D.) C18 reversed-phase HPLC. Column effluent was monitored by UV absorbance detection at 215 nm. The standard curve range was 1 to 250 ng/ml. The accuracy and precision values for the method were within +/- 10% and +/- 15%, respectively. A four-fold detectability enhancement was achieved using a 2.1 mm I.D. HPLC column relative to the more common 4.6 mm I.D. column. A performance comparison was made between the 2.1 mm I.D. column used for validation and a 4.6 mm I.D. column with the same stationary phase.
MeSH terms
Substances
LinkOut - more resources
Miscellaneous