A novel electric design for electromagnetic stimulation--the Slinky coil
- PMID: 7558066
- DOI: 10.1109/10.412658
A novel electric design for electromagnetic stimulation--the Slinky coil
Abstract
A novel coil design for inductive electromagnetic stimulation of neural cells has been simulated and experimentally tested. This coil improves the focal effect of a magnetic stimulator, and it reduces its inductance, hence the efficiency of the system is improved. The basic structure of the device is derived from the popular "Slinky" toy. The actual device is formed by winding different numbers of loops forming a helical coil on a half torus. The loops are bunched at the axis of the torus. The coil, due to its geometry, generates a unique distribution of eddy currents in nearby tissues which is favorable compared to a solenoid type stimulator. This renders the Slinky coil more selective than conventional coils used for magnetic stimulation. The distribution of eddy currents was analyzed using Matlab, following Faraday's Law of Induction. Improved focality permits the current through the coil to be reduced for the same effect. In addition, the reduced inductance of the Slinky coil decreases the power requirement; thus, the improved efficiency of the system may allow the generation of bursts of pulses, and expand the utilization of the system to possible functional activation of certain neuro-muscular structures when peripheral nerves are stimulated.
Similar articles
-
Cylindrical tissue model for magnetic field stimulation of neurons: effects of coil geometry.IEEE Trans Biomed Eng. 1995 Sep;42(9):934-41. doi: 10.1109/10.412660. IEEE Trans Biomed Eng. 1995. PMID: 7558068
-
Magnetic coil design considerations for functional magnetic stimulation.IEEE Trans Biomed Eng. 2000 May;47(5):600-10. doi: 10.1109/10.841332. IEEE Trans Biomed Eng. 2000. PMID: 10851804
-
Toroidal coil models for transcutaneous magnetic stimulation of nerves.IEEE Trans Biomed Eng. 2001 Apr;48(4):434-41. doi: 10.1109/10.915709. IEEE Trans Biomed Eng. 2001. PMID: 11322531
-
Technical and practical aspects of magnetic nerve stimulation.J Clin Neurophysiol. 1991 Jan;8(1):10-25. doi: 10.1097/00004691-199101000-00004. J Clin Neurophysiol. 1991. PMID: 2019644 Review.
-
[Apparatus and areas of application for low-frequency magnetotherapy].Med Tekh. 1974 May;3(0):41-6. Med Tekh. 1974. PMID: 4599934 Review. Russian. No abstract available.
Cited by
-
Electric field depth-focality tradeoff in transcranial magnetic stimulation: simulation comparison of 50 coil designs.Brain Stimul. 2013 Jan;6(1):1-13. doi: 10.1016/j.brs.2012.02.005. Epub 2012 Mar 21. Brain Stimul. 2013. PMID: 22483681 Free PMC article.
-
High-frequency rTMS over the left DLPFC improves the response inhibition control of young healthy participants: an ERP combined 1H-MRS study.Front Psychol. 2023 May 18;14:1144757. doi: 10.3389/fpsyg.2023.1144757. eCollection 2023. Front Psychol. 2023. PMID: 37275686 Free PMC article.
-
Effect of contour shape of nervous system electromagnetic stimulation coils on the induced electrical field distribution.Biomed Eng Online. 2002 May 14;1:1. doi: 10.1186/1475-925x-1-1. Biomed Eng Online. 2002. PMID: 12437767 Free PMC article.
-
Optimal Design of Array Coils for Multi-Target Adjustable Electromagnetic Brain Stimulation System.Bioengineering (Basel). 2023 May 9;10(5):568. doi: 10.3390/bioengineering10050568. Bioengineering (Basel). 2023. PMID: 37237638 Free PMC article.
-
Gyri-precise head model of transcranial direct current stimulation: improved spatial focality using a ring electrode versus conventional rectangular pad.Brain Stimul. 2009 Oct;2(4):201-7, 207.e1. doi: 10.1016/j.brs.2009.03.005. Brain Stimul. 2009. PMID: 20648973 Free PMC article.