An alternative limb lead system for electrocardiographs in emergency patients
- PMID: 7662053
- DOI: 10.1016/0735-6757(95)90160-4
An alternative limb lead system for electrocardiographs in emergency patients
Abstract
It is occasionally difficult to record the standard 12-lead electrocardiograph (ECG) in emergency patients. The aim of this study was to evaluate the influence on electrocardiographic wave form recordings of moving the location of electrodes from the standard limb lead position to the trunk. The participants were 10 normal subjects and 20 patients with heart disease. In the new lead system, the limb electrodes were placed on the anterior acromial region and the anterior superior iliac spine using adhesive electrodes. Conventional 12-lead ECGs were recorded by the standard and the new lead system simultaneously in the supine position. Wave form analysis was done by an automatic analysis program. Motion artifacts in the recordings were less in the new lead system. The R wave amplitude of the new lead system increased in leads II, III and aVF, and decreased in leads I and aVL. However, the amplitudes of each wave obtained by standard electrocardiography and the new lead system correlated well (y = 1.008x + 2.038, r = 0.99, n = 2,880). In 99.6% of all wave forms, the differences in amplitudes were within 5% of the values of standard recordings. The average of differences in the ST-segment was 2.6 +/- 11.4 microV. The frontal plane QRS axis obtained by the new lead system showed a vertical shift of 7.8 +/- 8.5 degrees (y = 0.94911x + 10.346, r = 0.98, n = 30). The recording errors produced by the new lead system were within the permissible range of variation. The new lead system is a reasonable alternative for recording ECGs if application of the standard lead is difficult in an emergency.
Comment in
-
Modified limb lead placement for rapid ECG.Am J Emerg Med. 1995 Sep;13(5):603-4. doi: 10.1016/0735-6757(95)90181-7. Am J Emerg Med. 1995. PMID: 7662065 No abstract available.
Similar articles
-
Effects of limb electrode placement on the 12- and 16-lead electrocardiogram.J Electrocardiol. 2008 Nov-Dec;41(6):536-45. doi: 10.1016/j.jelectrocard.2008.07.023. Epub 2008 Sep 13. J Electrocardiol. 2008. PMID: 18790505
-
On the normal scalar ECG. A new classification system considering age, sex and heart position.Acta Med Scand Suppl. 1984;691:1-147. Acta Med Scand Suppl. 1984. PMID: 6240193
-
Effects of electrocardiograph frequency filters on P-QRS-T amplitudes of the feline electrocardiogram.Am J Vet Res. 1995 Nov;56(11):1534-40. Am J Vet Res. 1995. PMID: 8585669
-
Potential solutions for providing standard electrocardiogram recordings from nonstandard recording sites.J Electrocardiol. 2008 May-Jun;41(3):207-10. doi: 10.1016/j.jelectrocard.2008.02.001. J Electrocardiol. 2008. PMID: 18433611 Review.
-
Technical mistakes during the acquisition of the electrocardiogram.Ann Noninvasive Electrocardiol. 2009 Oct;14(4):389-403. doi: 10.1111/j.1542-474X.2009.00328.x. Ann Noninvasive Electrocardiol. 2009. PMID: 19804517 Free PMC article. Review.
Cited by
-
Discovering substitutes to the conventional lead positions of three-lead electrocardiogram monitoring.Ann Card Anaesth. 2019 Apr-Jun;22(2):236-238. doi: 10.4103/aca.ACA_89_18. Ann Card Anaesth. 2019. PMID: 30971615 Free PMC article. No abstract available.
-
Phenomenal, bodily and brain correlates of fictional reappraisal as an implicit emotion regulation strategy.Cogn Affect Behav Neurosci. 2019 Aug;19(4):877-897. doi: 10.3758/s13415-018-00681-0. Cogn Affect Behav Neurosci. 2019. PMID: 30610654
-
Modified torso vs distal limb electrode placement for performing ECGs in children: A method comparison study.J Electrocardiol. 2024 Nov-Dec;87:153810. doi: 10.1016/j.jelectrocard.2024.153810. Epub 2024 Sep 24. J Electrocardiol. 2024. PMID: 39442283
-
Enhancing cognitive and motor performance through mental training: The interplay between temporal preparation, inhibition and autonomic arousal.Cogn Affect Behav Neurosci. 2025 Apr 30. doi: 10.3758/s13415-025-01301-4. Online ahead of print. Cogn Affect Behav Neurosci. 2025. PMID: 40304891
-
Modified limb lead ECG system effects on electrocardiographic wave amplitudes and frontal plane axis in sinus rhythm subjects.Anatol J Cardiol. 2017 Jan;17(1):46-54. doi: 10.14744/AnatolJCardiol.2016.6843. Epub 2016 Jun 29. Anatol J Cardiol. 2017. PMID: 27443474 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical