Predictors of spirometric test failure: a comparison of the 1983 and 1993 acceptability criteria from the European Community for Coal and Steel
- PMID: 7663642
- PMCID: PMC1128292
- DOI: 10.1136/oem.52.8.547
Predictors of spirometric test failure: a comparison of the 1983 and 1993 acceptability criteria from the European Community for Coal and Steel
Abstract
Objectives: To identify, in this general population study, predictors of spirometric test failure on the 1983 and 1993 acceptability criteria from the European Community for Coal and Steel (ECCS).
Methods: All men aged 30-46 years living in western Norway (n = 45,380) were invited to join a cross sectional community survey. Respiratory symptoms, smoking habits, and marital status were found from self administered questionnaires, and measurements of forced expiratory volume in one second (FEV1) and forced vital capacity (FVC) were performed with dry wedge bellow spirometers.
Results: Altogether 29,611 subjects (65%) participated in this survey. Spirometric recordings were obtained in 26,803 of these (91%) of whom 1.6% failed the 1983 criterion (< or = 300 ml between the two highest FVC values) and 9.5% failed the 1993 criteria (< or = 5% or 100 ml between the two highest FEV1 and the two highest FVC values). Spirometric failures on both criteria were more prevalent in never smokers, single men, and subjects with respiratory symptoms than in ever smokers, married, and asymptomatic subjects. Failure of the 1993 criteria increased with age and declining height. Morning cough and phlegm, breathlessness uphill, attacks of breathlessness, and wheezing were related to failure of the 1993 criteria after adjustment for demographic variables and smoking, whereas only breathlessness uphill was related to failure of the 1983 criterion.
Conclusions: In men aged 30-46 years, spirometric test failures on both the 1983 and 1993 ECCS acceptability criteria occurred more often in never smokers than in smokers and ex-smokers after adjustment for other covariables. Spirometric test failure with the 1993 criteria also varied with height and most respiratory symptoms. The higher failure rates found in non-smokers, in shorter, and in single men could be due to late compression of the airways, smaller lung volumes, and poor general health, respectively.
Similar articles
-
Association of years of occupational quartz exposure with spirometric airflow limitation in Norwegian men aged 30-46 years.Thorax. 1998 Aug;53(8):649-55. doi: 10.1136/thx.53.8.649. Thorax. 1998. PMID: 9828850 Free PMC article.
-
Forced expiratory volume in 1 second (FEV1) and forced vital capacity (FVC) variability in asymptomatic never-smoking men.Clin Physiol. 1998 Jul;18(4):387-96. doi: 10.1046/j.1365-2281.1998.00115.x. Clin Physiol. 1998. PMID: 9715766
-
[Spirometry in a population of coal miners in Paipa, Colombia].Biomedica. 2017 Dec 1;37(4):498-506. doi: 10.7705/biomedica.v37i4.3364. Biomedica. 2017. PMID: 29373770 Spanish.
-
Fixed ratio or lower limit of normal for the FEV1 /VC ratio: relation to symptoms and extended lung function tests.Clin Physiol Funct Imaging. 2017 May;37(3):263-269. doi: 10.1111/cpf.12294. Epub 2015 Oct 6. Clin Physiol Funct Imaging. 2017. PMID: 26443700
-
Relations between respiratory symptoms and spirometric values in young adults: the European community respiratory health study.Respir Med. 2004 Oct;98(10):1025-33. doi: 10.1016/j.rmed.2004.01.013. Respir Med. 2004. PMID: 15481280
Cited by
-
Spirometry quality predictors in a large multistate prospective study.Respir Med. 2021 Nov;188:106618. doi: 10.1016/j.rmed.2021.106618. Epub 2021 Sep 17. Respir Med. 2021. PMID: 34571455 Free PMC article.
-
Feasibility of measuring lung function in preschool children.Thorax. 2002 Dec;57(12):1021-7. doi: 10.1136/thorax.57.12.1021. Thorax. 2002. PMID: 12454295 Free PMC article.
-
Association of years of occupational quartz exposure with spirometric airflow limitation in Norwegian men aged 30-46 years.Thorax. 1998 Aug;53(8):649-55. doi: 10.1136/thx.53.8.649. Thorax. 1998. PMID: 9828850 Free PMC article.
-
Conformity of spirometric tests with acceptability criteria and assessment of confounding factors in routine clinical practice.Ann Thorac Med. 2025 Jul-Sep;20(3):176-182. doi: 10.4103/atm.atm_232_24. Epub 2025 Jun 16. Ann Thorac Med. 2025. PMID: 40786888 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical