Detection of hepatitis C virus infection with recombinant immunoblot assay, synthetic immunoblot assay, and polymerase chain reaction
- PMID: 7685380
- DOI: 10.1002/jcla.1860070306
Detection of hepatitis C virus infection with recombinant immunoblot assay, synthetic immunoblot assay, and polymerase chain reaction
Abstract
The newly developed immunoblot assay, RIBA SIA (recombinant and synthetic polypeptide immunoblot assay), Chiron, Calif., was compared with the commercially available second generation recombinant immunoblot assay (RIBA-2) for the detection of antibody to hepatitis C virus (anti-HCV). The two immunoblot tests were also compared with the polymerase chain reaction (PCR) for the detection of HCV RNA. Ninety-one percent of samples reactive by RIBA-2 were positive for anti-HCV by RIBA SIA. A total of 31% of RIBA-2 indeterminate samples became reactive by RIBA SIA, 24% became non-reactive, and 45% remained the same. Samples reactive by RIBA-2 or SIA from different risk groups, were mostly positive (67-100%) by PCR for HCV RNA. All indeterminate samples from hemophiliacs and intravenous drug users were PCR positive. RIBA SIA is more sensitive and specific than RIBA-2 and correlates well with PCR results.