Fretting corrosion accelerates crevice corrosion of modular hip tapers
- PMID: 7703534
- DOI: 10.1002/jab.770060104
Fretting corrosion accelerates crevice corrosion of modular hip tapers
Abstract
The use of multiple-component systems in orthopedic surgery gives the surgeon increased flexibility in choosing the optimal implant, but introduces the possibility of interfacial corrosion. Such corrosion could limit the longevity of prostheses due either to tissue reactions to corrosion products, or to device failure. The incidence and nature of corrosion of modular total hips was evaluated in a consecutive series of 79 retrieved implants from University Hospitals of Cleveland. Surfaces were examined with stereo- and scanning electron microscopy. Several laboratory studies were undertaken to examine mechanisms that might contribute to the initiation of corrosion. The first set of experiments investigated the effect of head neck extension; the second study looked at the effect of material combinations on fretting corrosion and crevice corrosion. Analysis of retrieved implants demonstrated that fretting corrosion played a major role in the initiation of interface corrosion, and that a correlation existed between corrosion and length of neck extensions. Laboratory studies showed that longer head neck extensions may be more susceptible to fretting corrosion because of an instability at the interface. Short-term mixed-metal corrosion studies demonstrated that the coupling of cobalt and titanium alloys did not render the interface more susceptible to corrosion. It is hypothesized that fretting corrosion contributes to the initiation of modular interface corrosion, and that the problem can be reduced by design changes that increase the stability of the interface.
Similar articles
-
An analysis of the head-neck taper interface in retrieved hip prostheses.Clin Orthop Relat Res. 1994 Mar;(300):162-7. Clin Orthop Relat Res. 1994. PMID: 8131330
-
In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling.J Biomed Mater Res. 1993 Dec;27(12):1533-44. doi: 10.1002/jbm.820271210. J Biomed Mater Res. 1993. PMID: 8113241
-
Mixed-metal fretting corrosion of Ti6Al4V and wrought cobalt alloy.J Biomed Mater Res. 1995 Jul;29(7):867-73. doi: 10.1002/jbm.820290712. J Biomed Mater Res. 1995. PMID: 7593026
-
Corrosion at the head-neck interface of current designs of modular femoral components: essential questions and answers relating to corrosion in modular head-neck junctions.Bone Joint J. 2016 May;98-B(5):579-84. doi: 10.1302/0301-620X.98B5.35592. Bone Joint J. 2016. PMID: 27143725 Review.
-
[Biodeterioration and corrosion of metallic implants and prostheses].Medicina (B Aires). 1993;53(3):260-74. Medicina (B Aires). 1993. PMID: 8114635 Review. Spanish.
Cited by
-
Mechanisms of Adverse Local Tissue Reactions to Hip Implants.Front Bioeng Biotechnol. 2019 Jul 30;7:176. doi: 10.3389/fbioe.2019.00176. eCollection 2019. Front Bioeng Biotechnol. 2019. PMID: 31417898 Free PMC article. Review.
-
Diagnosis and Management of Adverse Reactions to Metal Debris.Oper Tech Orthop. 2019 Sep;29(3):100732. doi: 10.1016/j.oto.2019.100732. Epub 2019 Oct 1. Oper Tech Orthop. 2019. PMID: 32296268 Free PMC article.
-
Clinical and radiological outcomes and analysis of failures of modular revisions stems at long-term follow-up: a systematic review and meta-analysis.Ann Jt. 2023 Nov 25;9:1. doi: 10.21037/aoj-23-32. eCollection 2024. Ann Jt. 2023. PMID: 38529295 Free PMC article.
-
In Vivo Damage of the Head-Neck Junction in Hard-on-Hard Total Hip Replacements: Effect of Femoral Head Size, Metal Combination, and 12/14 Taper Design.Materials (Basel). 2017 Jul 1;10(7):733. doi: 10.3390/ma10070733. Materials (Basel). 2017. PMID: 28773095 Free PMC article.
-
Comparison of ASTM F2129 and ASTM F746 for Evaluating Crevice Corrosion.J Test Eval. 2019 Jul;47(4):2497-2511. doi: 10.1520/jte20180585. Epub 2019 Feb 19. J Test Eval. 2019. PMID: 37680964 Free PMC article.