Effect of thromboplastin and coagulometer interaction on the precision of the International Normalised Ratio
- PMID: 7706513
- PMCID: PMC502253
- DOI: 10.1136/jcp.48.1.13
Effect of thromboplastin and coagulometer interaction on the precision of the International Normalised Ratio
Abstract
Aims: To examine the magnitude of thromboplastin and coagulometer interactions on the precision of International Normalised Ratio (INR) values when the manufacturers' recommended instrument specific International Sensitivity Index (ISI) values are adopted for the INR calculation.
Methods: The variability of INR values obtained from four automated phototopical coagulometers frequently used in North American laboratories was studied. When used with five commercial thromboplastins of moderate to high sensitivity (ISI values 0.92-1.97), 20 prothrombin time results were generated for each of a population of 98 patients on established warfarin treatment.
Results: The mean INR values of the patients ranged from 2.05 to 2.81, depending on which reagent/coagulometer combination was used. Within patient variation increased as the median INR value increased. The mean coefficient of variation of within patient INR values was 10%; the mean coefficient of variation of the prothrombin time results in seconds and prothrombin time ratio were 21 and 18%, respectively.
Conclusions: There was considerable bias in the estimated ISI values of the thromboplastins compared with the manufacturers' instrument specific ISI value. Despite this apparent imperfection, our study clearly showed that the INR is preferable to other prothrombin time reporting formats for assessing the degree of anticoagulation for patients on warfarin treatment.
Similar articles
-
The value of plasma calibrants in correcting coagulometer effects on international normalized ratios. An international multicenter study.Am J Clin Pathol. 1995 Mar;103(3):358-65. doi: 10.1093/ajcp/103.3.358. Am J Clin Pathol. 1995. PMID: 7872261
-
A comparison of five commercial thromboplastins: ISI re-evaluation on an automated coagulometer.Clin Lab Haematol. 1995 Mar;17(1):41-5. doi: 10.1111/j.1365-2257.1995.tb00315.x. Clin Lab Haematol. 1995. PMID: 7621627
-
The European Concerted Action on Anticoagulation (ECAA): field studies of coagulometer effects on the ISI of ECAA thromboplastins.Thromb Haemost. 1998 Oct;80(4):615-23. Thromb Haemost. 1998. PMID: 9798981
-
Standardization of Prothrombin Time/International Normalized Ratio (PT/INR).Int J Lab Hematol. 2021 Feb;43(1):21-28. doi: 10.1111/ijlh.13349. Epub 2020 Sep 26. Int J Lab Hematol. 2021. PMID: 32979036 Review.
-
[Prothrombin time and its standardization].Rinsho Byori. 2002 Aug;50(8):779-85. Rinsho Byori. 2002. PMID: 12373814 Review. Japanese.
Cited by
-
Accuracy and precision of the TAS analyser for near-patient INR testing by non-pathology staff in the community.J Clin Pathol. 1998 Jan;51(1):68-72. doi: 10.1136/jcp.51.1.68. J Clin Pathol. 1998. PMID: 9577376 Free PMC article.
-
Clinical utilization of the international normalized ratio (INR).J Clin Lab Anal. 2000;14(3):101-14. doi: 10.1002/(sici)1098-2825(2000)14:3<101::aid-jcla4>3.0.co;2-a. J Clin Lab Anal. 2000. PMID: 10797608 Free PMC article. Review.
-
Adaptive force sonorheometry for assessment of whole blood coagulation.Clin Chim Acta. 2010 May 2;411(9-10):638-44. doi: 10.1016/j.cca.2010.01.018. Epub 2010 Jan 22. Clin Chim Acta. 2010. PMID: 20096680 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources