Effect of respiration on Doppler parameters of normal tricuspid porcine bioprosthetic valves
- PMID: 7730682
- DOI: 10.1007/BF01148954
Effect of respiration on Doppler parameters of normal tricuspid porcine bioprosthetic valves
Abstract
The Doppler indexes of tricuspid porcine bioprosthetic valves were evaluated in twelve patients without clinical and two-dimensional echocardiographic evidence of valve dysfunction. Peak and mean pressure gradients across the prostheses were measured using the simplified Bernoulli equation. All the Doppler measurements were compared during inspiration and expiration. During inspiration peak velocity, peak gradient and mean gradient (1.52 +/- 0.28 m/s; 9.7 +/- 3.05 mmHg; 4.07 +/- 1.16 mmHg) were significantly higher than during expiration (1.28 +/- 0.8 m/s; 6.58 +/- 2.7 mmHg; 2.98 +/- 1.13 mmHg; p < 0.01) but pressure half time was not significantly different (122 +/- 62 ms versus 134 +/- 75 ms; p > 0.05). Inspiratory range of peak velocities, peak gradients, mean gradients and pressure half times were respectively 0.8-2.04 m/s; 4.9-16.6 mmHg; 1.2-7.2 mmHg; 42-340 ms while expiratory range of values was 0.8-1.93 m/s; 2.6-15 mmHg; 1.1-5.7 mmHg; 46-345 ms. These data suggest that even very long pressure half times do not indicate valve dysfunction. This study demonstrates that large variation of Doppler parameters are present during respiration and could produce inaccuracy in the assessment of bioprostheses in tricuspid position if they are not taken in consideration.
Similar articles
-
Doppler echocardiographic evaluation of St. Jude Medical valves in the tricuspid position.J Heart Valve Dis. 1993 May;2(3):279-86. J Heart Valve Dis. 1993. PMID: 8269120
-
Transthoracic echocardiographic Doppler metrics in evaluating bioprosthetic tricuspid valve dysfunction.Echocardiography. 2024 May;41(5):e15835. doi: 10.1111/echo.15835. Echocardiography. 2024. PMID: 38784978
-
[Doppler echocardiography in assessing mechanical and biological heart valve prostheses].G Ital Cardiol. 1988 Feb;18(2):121-34. G Ital Cardiol. 1988. PMID: 3410201 Italian.
-
[Evaluation of the normal bioprosthetic Intact aortic valve by Doppler echocardiography].Arch Mal Coeur Vaiss. 1990 Dec;83(14):2039-44. Arch Mal Coeur Vaiss. 1990. PMID: 2126711 Review. French.
-
Doppler echocardiography assessment of prosthetic heart valves.Echocardiography. 1990 Mar;7(2):97-114. doi: 10.1111/j.1540-8175.1990.tb00353.x. Echocardiography. 1990. PMID: 10149195 Review.