Comparing two heat and moisture exchangers with one vaporizing humidifier in patients with minute ventilation greater than 10 L/min
- PMID: 7750340
- DOI: 10.1378/chest.107.5.1411
Comparing two heat and moisture exchangers with one vaporizing humidifier in patients with minute ventilation greater than 10 L/min
Abstract
Study objective: To evaluate in patients submitted to minute ventilation > 10 L/min the ability to preserve patients' heat and humidity of two heat and moisture exchangers (HMEs) and one vaporizing humidifier (VH).
Design: Prospective, randomized, comparative, non-blinded study.
Setting: Intensive care unit of a university hospital.
Patients: Nine tracheally intubated, mechanically ventilated patients, sedated and submitted to mechanical ventilation with minute ventilation > 10 L/min.
Interventions: Using the psychrometric method, relative humidity (RH) and absolute humidity (AH) of inspired gas were obtained as well as temperature of inspired gas and tracheal temperatures (maximal and minimal). Following a randomized order, each patient was ventilated for two 24-h periods with a vaporizing humidifier (Bennett Cascade 2, Bennett; France) and one of two HMEs: Pall Ultipor filter BB50 (Pall Biomedical; France) or DAR Hygroster filter (Peters; France). Both were first tested for a 45-min period and then the HME that achieved the best performance in terms of temperature and water preservation was tested for 24 h.
Measurements and results: During the 45-min test period, the Pall Ultipor HME achieved a lower performance than the other two systems for any of the studied parameters (p < 0.05 to p < 0.0001). The DAR Hygroster HME achieved lower temperature of inspired gas (29.9 vs 32.0 degrees C, p < 0.005) and lower absolute humidity (29.3 vs 33.2 mg H2O/L, p < 0.005) than the Bennett Cascade 2. After 24 h of use, lower values of temperature of inspired gas (28.5 vs 32.0 degrees C, p < 0.002) and of AH (28.0 vs 33.6 mg H2O/L, p < 0.001) were obtained with the DAR Hygroster HME than with the Bennett Cascade 2. No differences were found between the two systems for the other tested parameters. At that time, no patients had RH lower than 97% and absolute humidity lower than 23 mg H2O/L with the use of the DAR Hygroster HME.
Conclusions: In patients with minute ventilation > 10 L/min, the DAR Hygroster HME showed a thermic and humidification capability similar to the reference system, the Bennett Cascade 2 VH. In these patients, the Pall Ultipor HME had a significantly lower capability.
Similar articles
-
Preservation of humidity and heat of respiratory gases in patients with a minute ventilation greater than 10 L/min.Crit Care Med. 1994 Nov;22(11):1871-6. Crit Care Med. 1994. PMID: 7956294 Clinical Trial.
-
Performance evaluation of three vaporizing humidifiers and two heat and moisture exchangers in patients with minute ventilation > 10 L/min.Chest. 1992 Nov;102(5):1347-50. doi: 10.1378/chest.102.5.1347. Chest. 1992. PMID: 1424849 Clinical Trial.
-
Changing a hydrophobic heat and moisture exchanger after 48 hours rather than 24 hours: a clinical and microbiological evaluation.Intensive Care Med. 1999 Nov;25(11):1237-43. doi: 10.1007/s001340051051. Intensive Care Med. 1999. PMID: 10654207
-
Humidification for ventilated patients.Intensive Crit Care Nurs. 1992 Mar;8(1):2-9. doi: 10.1016/0964-3397(92)90003-3. Intensive Crit Care Nurs. 1992. PMID: 1559020 Review.
-
Respiratory gas conditioning in infants with an artificial airway.Semin Neonatol. 2002 Oct;7(5):369-77. doi: 10.1053/siny.2002.0131. Semin Neonatol. 2002. PMID: 12464499 Review.
Cited by
-
Heat and Moisture Exchangers and Humidification Efficacy in Pediatrics: Effects of Ventilator Settings and ETT Leakage.Crit Care Res Pract. 2012;2012:585280. doi: 10.1155/2012/585280. Epub 2012 Aug 1. Crit Care Res Pract. 2012. PMID: 22900156 Free PMC article. No abstract available.
-
Importance and interpretation of fast-response airway hygrometry during ventilation of anesthetized patients.J Clin Monit Comput. 2007 Jun;21(3):137-46. doi: 10.1007/s10877-006-9065-5. Epub 2007 Mar 16. J Clin Monit Comput. 2007. PMID: 17364215
-
Heat and moisture exchangers versus heated humidifiers for mechanically ventilated adults and children.Cochrane Database Syst Rev. 2017 Sep 14;9(9):CD004711. doi: 10.1002/14651858.CD004711.pub3. Cochrane Database Syst Rev. 2017. PMID: 28905374 Free PMC article.
-
In vitro and in vivo evaluation of a new active heat moisture exchanger.Crit Care. 2004 Oct;8(5):R281-8. doi: 10.1186/cc2904. Epub 2004 Jun 28. Crit Care. 2004. PMID: 15469569 Free PMC article.
-
Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes.Crit Care Res Pract. 2018 Dec 10;2018:3258396. doi: 10.1155/2018/3258396. eCollection 2018. Crit Care Res Pract. 2018. PMID: 30652032 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical