MR imaging of the abdomen with a phased-array multicoil: prospective clinical evaluation
- PMID: 7754009
- DOI: 10.1148/radiology.195.3.7754009
MR imaging of the abdomen with a phased-array multicoil: prospective clinical evaluation
Abstract
Purpose: To prospectively compare use of a phased-array multicoil and a conventional body coil in abdominal MR imaging.
Materials and methods: Thirteen patients (seven men, six women; mean age, 55 years) underwent imaging with a phased-array multicoil and with a conventional body coil. Four pulse sequences were used: T2-weighted spin echo (SE), magnetization-prepared gradient-recalled echo (GRE), breath-hold fast SE, and echo planar (EP).
Results: Lesion detection improved the most on fast SE, multicoil-acquired images. Signal-to-noise ratio (S/N) increased 64% with fast SE (P = .0005) and EP (P < .0109) sequences. Contrast-to-noise ratio (C/N) doubled (P < .05) with T2-weighted SE sequences. Lesion conspicuity improved on multicoil-acquired images with all fast sequences (magnetization-prepared GRE, P = .015; fast SE, P = .002; EP imaging, P = .013). There was little difference in respiratory and vascular artifact. Depiction of most abdominal structures improved (P < .01).
Conclusion: Use of the phased-array multicoil provides better MR images of the abdomen than does use of a conventional body coil.
Similar articles
-
Fast spin-echo MR images of the pelvis obtained with a phased-array coil: value in localizing and staging prostatic carcinoma.AJR Am J Roentgenol. 1993 Sep;161(3):601-6. doi: 10.2214/ajr.161.3.8352116. AJR Am J Roentgenol. 1993. PMID: 8352116
-
Liver T2-weighted MR imaging: comparison of fast and conventional half-Fourier single-shot turbo spin-echo, breath-hold turbo spin-echo, and respiratory-triggered turbo spin-echo sequences.Radiology. 1997 Jun;203(3):766-72. doi: 10.1148/radiology.203.3.9169702. Radiology. 1997. PMID: 9169702
-
[Phased-array superficial coil and breath holding technique in MRI of the liver. Comparison of conventional spin echo sequences with rapid fat suppressing gradient echo and turbo-spin sequences].Radiologe. 1995 Dec;35(12):919-24. Radiologe. 1995. PMID: 8584635 German.
-
[Fast and ultra-fast sequences in MRI of the pelvis and retroperitoneum].Radiologe. 1995 Dec;35(12):936-44. Radiologe. 1995. PMID: 8584637 Review. German.
-
Technologic advances in abdominal MR imaging.Radiology. 2001 Aug;220(2):310-20. doi: 10.1148/radiology.220.2.r01au22310. Radiology. 2001. PMID: 11477231 Review.
Cited by
-
Signal-to-noise ratio and parallel imaging performance of commercially available phased array coils in 3.0 T brain magnetic resonance imaging.Radiol Phys Technol. 2015 Jul;8(2):305-11. doi: 10.1007/s12194-015-0321-6. Epub 2015 May 26. Radiol Phys Technol. 2015. PMID: 26009490
-
MRI of gastric carcinoma: results of T and N-staging in an in vitro study.World J Gastroenterol. 2009 Aug 28;15(32):3992-8. doi: 10.3748/wjg.15.3992. World J Gastroenterol. 2009. PMID: 19705493 Free PMC article.
-
T2-weighted breathold imaging of the liver: a quantitative and qualitative comparison of fast spin echo and half Fourier single shot fast spin echo imaging.MAGMA. 1999 Oct;9(1-2):42-51. doi: 10.1007/BF02634591. MAGMA. 1999. PMID: 10555172
-
High-resolution MRI of pancreatic masses with a new circularly polarized body phased-array coil.Eur Radiol. 1997;7(5):643-8. doi: 10.1007/BF02742917. Eur Radiol. 1997. PMID: 9166559
-
Application of high-density 2D receiver coil arrays for improved SNR in prostate MRI.Magn Reson Med. 2025 Feb;93(2):850-863. doi: 10.1002/mrm.30289. Epub 2024 Sep 25. Magn Reson Med. 2025. PMID: 39322985