Infection rates in single- and double-lumen central venous catheters in critically ill patients
- PMID: 7780350
Infection rates in single- and double-lumen central venous catheters in critically ill patients
Abstract
Background: Catheter-related sepsis is a well known complication in critically ill patients receiving total parenteral nutrition. Micro-organisms may travel from the skin puncture wound along the external surface of the catheter or from the hub through the lumen of the catheter, to be shed into the circulation causing bacteraemia and sepsis. The incidence of sepsis is said to be about three times greater with multiple-lumen catheters than with single-lumen catheters.
Methods: Fifty patients admitted to our intensive care unit were randomly assigned to have either a single-lumen (Group I, 25 patients) or double-lumen (Group II, 25 patients) hydromer-coated, polyurethane central venous catheters of the same gauge and length, placed through the infraclavicular subclavian approach using the Seldinger technique. A sterile dry gauze dressing at the skin puncture site was changed every day. Skin swabs were taken for semi-quantitative culture studies before catheter insertion and removal, from the proximal 3 cm (lying subcutaneously) and distal 5 cm of the catheter segments on removal of the catheter, and venous blood was drawn from a peripheral vein before removal of the catheter.
Results: The most common isolate was coagulase negative Staphylococcus. No significant difference was found between the two groups in the incidence of catheter-related infection and catheter-related sepsis.
Conclusion: Double-lumen central venous catheters placed in critically ill patients do not cause more sepsis than single-lumen catheters.
Similar articles
-
Triple- vs single-lumen central venous catheters. A prospective study in a critically ill population.Arch Intern Med. 1989 May;149(5):1139-43. Arch Intern Med. 1989. PMID: 2497712
-
Failure of antiseptic bonding to prevent central venous catheter-related infection and sepsis.Am Surg. 1996 Aug;62(8):641-6. Am Surg. 1996. PMID: 8712561 Clinical Trial.
-
Comparison of silver-impregnated with standard multi-lumen central venous catheters in critically ill patients.Crit Care Med. 2007 Apr;35(4):1032-9. doi: 10.1097/01.CCM.0000259378.53166.1B. Crit Care Med. 2007. PMID: 17334256 Clinical Trial.
-
Intravenous catheter-related infections.Adv Pediatr Infect Dis. 1995;10:337-68. Adv Pediatr Infect Dis. 1995. PMID: 7718211 Review.
-
Recommendations and reports about central venous catheter-related infection.Surg Infect (Larchmt). 2006;7 Suppl 2:S65-7. doi: 10.1089/sur.2006.7.s2-65. Surg Infect (Larchmt). 2006. PMID: 16895510 Review.
Cited by
-
The incidence of complications of central venous catheters at an intensive care unit.Ann Thorac Med. 2007 Apr;2(2):61-3. doi: 10.4103/1817-1737.32232. Ann Thorac Med. 2007. PMID: 19727348 Free PMC article.
-
Intravascular catheter related infections in children admitted on the paediatric wards of Mulago Hospital, Uganda.Afr Health Sci. 2008 Dec;8(4):206-16. Afr Health Sci. 2008. PMID: 20589126 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical