Intravascular membrane oxygenation and carbon dioxide removal with IVOX: can improved design and permissive hypercapnia achieve adequate respiratory support during severe respiratory failure?
- PMID: 7864733
- DOI: 10.1111/j.1525-1594.1994.tb03332.x
Intravascular membrane oxygenation and carbon dioxide removal with IVOX: can improved design and permissive hypercapnia achieve adequate respiratory support during severe respiratory failure?
Abstract
The intravenacaval oxygenator and carbon dioxide removal device (IVOX) conceived by Mortensen at CardioPulmonics is a diffusion-limited device capable of removing 30% of CO2 production of an adult at normocapnia with minimal reduction in ventilator requirements. Through mathematical modeling, an ex vivo venovenous bypass circuit to model the vena cava and animal models of severe smoke inhalation injury, the practice of permissive hypercapnia has been established to enhance CO2 removal by IVOX. By allowing the blood PCO2 to rise gradually, the CO2 excretion by IVOX can be linearly increased in a 1:1 relationship. Experimental and clinical studies have shown that CO2 removal by IVOX increased from 30-40 ml/min at normal blood PCO2 to 80-90 ml/min at PCO2 of 90 mm Hg. In addition, IVOX with permissive hypercapnia allowed a significant reduction in minute ventilation and peak airway pressure. Design changes could also improve the performance of IVOX. Increased surface area and mixing with more fibers and crimping in new prototypes of IVOX significantly increased CO2 removal and oxygen transfer. Active mixing in the blood to decrease the boundary layer resistance can further enhance gas exchange of IVOX. In conclusion, gas exchange by the current design of IVOX is limited, and improvements in design are needed for it to become a more clinically applicable device. Permissive hypercapnia can significantly enhance CO2 removal by IVOX as well as significantly reduce ventilator requirements.
Similar articles
-
Intravascular membrane oxygenator and carbon dioxide removal devices: a review of performance and improvements.ASAIO J. 1999 Jan-Feb;45(1):41-6. doi: 10.1097/00002480-199901000-00010. ASAIO J. 1999. PMID: 9952005 Review.
-
Stragegies to reduce surface area requirements for carbon dioxide removal for an intravenacaval gas exchange device.ASAIO J. 1995 Jul-Sep;41(3):M567-72. doi: 10.1097/00002480-199507000-00075. ASAIO J. 1995. PMID: 8573869
-
In vivo gas transfer performance of the intravascular oxygenator in acute respiratory failure.Artif Organs. 1994 Nov;18(11):840-5. doi: 10.1111/j.1525-1594.1994.tb03333.x. Artif Organs. 1994. PMID: 7864734
-
Intracorporeal CO2 removal and permissive hypercapnia to reduce airway pressure in acute respiratory failure. The theoretical basis for permissive hypercapnia with IVOX.ASAIO J. 1993 Apr-Jun;39(2):97-102. ASAIO J. 1993. PMID: 8324269
-
Intravascular membrane oxygenation and carbon dioxide removal--a new application for permissive hypercapnia?Thorac Cardiovasc Surg. 1992 Jun;40(3):115-20. doi: 10.1055/s-2007-1020127. Thorac Cardiovasc Surg. 1992. PMID: 1412375 Review.
Cited by
-
Bioengineering Progress in Lung Assist Devices.Bioengineering (Basel). 2021 Jun 28;8(7):89. doi: 10.3390/bioengineering8070089. Bioengineering (Basel). 2021. PMID: 34203316 Free PMC article. Review.
-
The Roles of Membrane Technology in Artificial Organs: Current Challenges and Perspectives.Membranes (Basel). 2021 Mar 28;11(4):239. doi: 10.3390/membranes11040239. Membranes (Basel). 2021. PMID: 33800659 Free PMC article. Review.
-
Intracorporeal gas exchange: current status and future development.Intensive Care Med. 1996 Feb;22(2):91-3. doi: 10.1007/BF01720713. Intensive Care Med. 1996. PMID: 8857114 No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical