The value of plasma calibrants in correcting coagulometer effects on international normalized ratios. An international multicenter study
- PMID: 7872261
- DOI: 10.1093/ajcp/103.3.358
The value of plasma calibrants in correcting coagulometer effects on international normalized ratios. An international multicenter study
Abstract
A study of lyophilized plasma calibrants in correcting for coagulometer effects on International Normalized Ratios (INR) has been conducted in an international survey. Prothrombin times were performed with the calibrants on 2 "common" thromboplastins and in-house thromboplastins in 3 brands of coagulometer at 37 centers. The International Sensitivity Indices (ISI) of 95 local systems with the calibrants were derived. The "true" INR of 10 test plasmas from warfarin-treated patients were established centrally using thromboplastin International Reference Preparations (IRP). International Normalized Ratios of these plasmas for each center were calculated using both the manufacturers' stated and the local ISI. The mean deviation of the 95 systems from the "true" INR of the warfarin plasmas was +14.4% with the manufacturers' ISI, but reduced to +1.04% with the local ISI. Local ISI determination with the calibrants avoids many of the difficulties of conventional thromboplastin calibrations. Plasmas from patients on warfarin and parallel manual PT with thromboplastin IRP are not required.
Similar articles
-
A comparison of lyophilized artificially depleted plasmas and lyophilized plasmas from patients receiving warfarin in correcting for coagulometer effects on international normalized ratios.Am J Clin Pathol. 1995 Mar;103(3):366-71. doi: 10.1093/ajcp/103.3.366. Am J Clin Pathol. 1995. PMID: 7872262
-
Standardization of coagulation tests.Southeast Asian J Trop Med Public Health. 1999;30 Suppl 3:79-85. Southeast Asian J Trop Med Public Health. 1999. PMID: 10926265
-
The correction of coagulometer effects on international normalized ratios: a multicentre evaluation.Br J Haematol. 1994 Jan;86(1):112-7. doi: 10.1111/j.1365-2141.1994.tb03260.x. Br J Haematol. 1994. PMID: 8011518
-
Precision and accuracy of the international normalized ratio in oral anticoagulant control.Haemostasis. 1996 Oct;26 Suppl 4:248-65. doi: 10.1159/000217306. Haemostasis. 1996. PMID: 8979131 Review.
-
Enhanced standardization of the International Normalized Ratio through the use of plasma calibrants: a concise review.Blood Coagul Fibrinolysis. 2000 Oct;11(7):583-90. doi: 10.1097/00001721-200010000-00001. Blood Coagul Fibrinolysis. 2000. PMID: 11085277 Review.
Cited by
-
Erratum to: Pharmacology of anticoagulants used in the treatment of venous thromboembolism.J Thromb Thrombolysis. 2016 Aug;42(2):296-311. doi: 10.1007/s11239-016-1363-2. J Thromb Thrombolysis. 2016. PMID: 27145758 Free PMC article.
-
The effects of freeze drying and freeze drying additives on the prothrombin time and the international sensitivity index.J Clin Pathol. 1999 Oct;52(10):744-8. doi: 10.1136/jcp.52.10.744. J Clin Pathol. 1999. PMID: 10674031 Free PMC article.
-
Accuracy and precision of the TAS analyser for near-patient INR testing by non-pathology staff in the community.J Clin Pathol. 1998 Jan;51(1):68-72. doi: 10.1136/jcp.51.1.68. J Clin Pathol. 1998. PMID: 9577376 Free PMC article.
-
Screening INR deviation of local prothrombin time systems.J Clin Pathol. 1998 May;51(5):356-9. doi: 10.1136/jcp.51.5.356. J Clin Pathol. 1998. PMID: 9708200 Free PMC article.
-
Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents.Biochem Med (Zagreb). 2012;22(1):121-6. doi: 10.11613/bm.2012.014. Biochem Med (Zagreb). 2012. PMID: 22384527 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources