. 1995 Mar;109(3):409-17; discussion 417-8.
doi: 10.1016/S0022-5223(95)70271-7.

Hemodynamic and physiologic changes during support with an implantable left ventricular assist device

Affiliations
Free article

Hemodynamic and physiologic changes during support with an implantable left ventricular assist device

P M McCarthy et al. J Thorac Cardiovasc Surg. 1995 Mar.
Free article

Abstract

To evaluate hemodynamic effectiveness and physiologic changes on the HeartMate 1000 IP left ventricular assist device (Thermo Cardiosystems, Inc., Woburn, Mass.), we studied 25 patients undergoing bridge to heart transplantation (35 to 63 years old, mean 50 years). All were receiving inotropic agents before left ventricular assist device implantation, 21 (84%) were supported with a balloon pump, and 7 (28%) were supported by extracorporeal membrane oxygenation. Six patients died, primarily of right ventricular dysfunction and multiple organ failure. Nineteen (76%) were rehabilitated, received a donor heart, and were discharged (100% survival after transplantation). Pretransplantation duration of support averaged 76 days (22 to 153 days). No thromboembolic events occurred in more than 1500 patient-days of support with only antiplatelet medications. Significant hemodynamic improvement was measured (before implantation to before explantation) in cardiac index (1.7 +/- 0.3 to 3.1 +/- 0.8 L/min per square meter; p < 0.001), left atrial pressure (23.7 +/- 7 to 9 +/- 7.5 mm Hg; p < 0.001), pulmonary artery pressure, pulmonary vascular resistance, and right ventricular volumes and ejection fraction. Both creatinine and blood urea nitrogen levels were significantly higher before implantation in patients who died while receiving support. Renal and liver function returned to normal before transplantation. We conclude that support with the HeartMate device improved hemodynamic and subsystem function before transplantation. Long-term support with the HeartMate device has a low risk of thromboemboli and makes a clinical trial of a portable HeartMate device a realistic alternative to medical therapy.