Reflotron cholesterol measurement in general practice: accuracy and detection of errors
- PMID: 7880074
- DOI: 10.1177/000456329403100605
Reflotron cholesterol measurement in general practice: accuracy and detection of errors
Abstract
Comparison of cholesterol determinations by nurses using a Reflotron analyser in a general practice setting showed a good correlation with plasma cholesterol determinations by wet chemistry in a clinical biochemistry laboratory. A limited number of comparisons did, however, give a much lower result on the Reflotron. In an experimental situation, small sample volumes (which could result from poor technique) were shown to produce falsely low readings. A simple method which may immediately detect falsely low Reflotron readings is discussed.
Similar articles
-
[Measurement of blood cholesterol, decentralized, using a Reflotron, or centralized in a laboratory. A comparison].Ugeskr Laeger. 1990 Nov 5;152(45):3346-8. Ugeskr Laeger. 1990. PMID: 2238225 Danish.
-
How accurately does the Reflotron dry-chemistry system measure plasma total cholesterol levels when used a community-screening device.Med J Aust. 1988 Aug 1;149(3):122-5. Med J Aust. 1988. PMID: 3398798
-
Reliability of the Reflotron in the determination of cholesterol.Scand J Clin Lab Invest. 1990 May;50(3):261-71. doi: 10.3109/00365519009091577. Scand J Clin Lab Invest. 1990. PMID: 2353155
-
A review of the analytic performance of the Reflotron System for cholesterol testing.Clin Ther. 1990 May-Jun;12(3):281-6. Clin Ther. 1990. PMID: 2199060 Review.
-
Improving the quality of plasma cholesterol measurements in primary care.Scand J Clin Lab Invest Suppl. 1990;198:43-8. Scand J Clin Lab Invest Suppl. 1990. PMID: 2189209 Review.
Cited by
-
Cardiovascular risk screening program in Australian community pharmacies.Pharm World Sci. 2010 Jun;32(3):373-80. doi: 10.1007/s11096-010-9379-8. Epub 2010 Mar 10. Pharm World Sci. 2010. PMID: 20217476
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical