On-line measurement of brain glutamate with an enzyme/polymer-coated tubular electrode
- PMID: 7915089
- DOI: 10.1021/ac00085a016
On-line measurement of brain glutamate with an enzyme/polymer-coated tubular electrode
Abstract
An on-line tubular electrode system is described which provides real time measurements of glutamate in brain microdialysate. It is based on the enzyme glutamate oxidase (EC 1.4.3.11) and the detection of H2O2 on a platinum electrode at 600 mV vs Ag/AgCl. The enzyme is immobilized in a layer of o-phenylenediamine electropolymerized at 750 mV in a phosphate buffer, pH 7.0. The layer is 10-15 nm thick and enables good enzyme loading and fast response time. The ability of the polymer to block out electroactive interferents like ascorbate and uric acid combined with a preoxidation system run at 600 mV resulted in a virtually interference free glutamate assay with a lower detection limit of 0.3 microM in the presence of physiological levels of interferents, a sensitivity of 4 nA/microM, and a linear region up to 30 microM. The system is optimized for use with a microdialysis probe implanted in the brain and perfused at 2 microL/min. It provides a time resolution of < 1 min and has been tested in vivo.
Similar articles
-
Functionalized conducting polymer as an enzyme-immobilizing substrate: an amperometric glutamate microbiosensor for in vivo measurements.Anal Chem. 2005 Aug 1;77(15):4854-60. doi: 10.1021/ac050558v. Anal Chem. 2005. PMID: 16053298
-
Extracellular glutamate: on-line monitoring using microdialysis coupled to enzyme-amperometric analysis.J Neurosci Methods. 1995 Aug;60(1-2):1-9. doi: 10.1016/0165-0270(94)00214-2. J Neurosci Methods. 1995. PMID: 8544467
-
Enzyme packed bed system for the on-line measurement of glucose, glutamate, and lactate in brain microdialysate.Anal Chem. 1992 Sep 1;64(17):1790-4. doi: 10.1021/ac00041a010. Anal Chem. 1992. PMID: 1416036
-
Continuous on-line measurement of cerebral hydrogen peroxide using enzyme-modified ring-disk plastic carbon film electrode.Anal Chem. 2002 Aug 1;74(15):3684-9. doi: 10.1021/ac011261+. Anal Chem. 2002. PMID: 12175154
-
Micro-flow in vivo analysis of L-glutamate with an on-line enzyme amplifier based on substrate recycling.Anal Sci. 2001 Jun;17(6):703-8. doi: 10.2116/analsci.17.703. Anal Sci. 2001. PMID: 11707939
Cited by
-
Microdialysis and microperfusion electrodes in neurologic disease monitoring.Fluids Barriers CNS. 2021 Dec 1;18(1):52. doi: 10.1186/s12987-021-00292-x. Fluids Barriers CNS. 2021. PMID: 34852829 Free PMC article. Review.
-
The mechanisms controlling physiologically stimulated changes in rat brain glucose and lactate: a microdialysis study.J Physiol. 1996 Oct 1;496 ( Pt 1)(Pt 1):49-57. doi: 10.1113/jphysiol.1996.sp021664. J Physiol. 1996. PMID: 8910195 Free PMC article.
-
Delayed production of adenosine underlies temporal modulation of swimming in frog embryo.J Physiol. 1998 Aug 15;511 ( Pt 1)(Pt 1):265-72. doi: 10.1111/j.1469-7793.1998.265bi.x. J Physiol. 1998. PMID: 9679180 Free PMC article.
-
Recent trends in microdialysis sampling integrated with conventional and microanalytical systems for monitoring biological events: a review.Anal Chim Acta. 2009 Sep 28;651(1):1-14. doi: 10.1016/j.aca.2009.07.064. Epub 2009 Aug 3. Anal Chim Acta. 2009. PMID: 19733728 Free PMC article. Review.
-
Microsensors for in vivo Measurement of Glutamate in Brain Tissue.Sensors (Basel). 2008 Nov 4;8(11):6860-6884. doi: 10.3390/s8116860. Sensors (Basel). 2008. PMID: 27873904 Free PMC article. Review.