Intracardiac echocardiography during radiofrequency catheter ablation of cardiac arrhythmias in humans
- PMID: 7930260
- DOI: 10.1016/0735-1097(94)90119-8
Intracardiac echocardiography during radiofrequency catheter ablation of cardiac arrhythmias in humans
Abstract
Objectives: The purpose of this study was to describe our preliminary experience using catheter-based intracardiac echocardiography as an adjunct to biplane fluoroscopy for guiding radiofrequency catheter ablation of atrial arrhythmias in the right side of the heart.
Background: Catheter ablation requires precise positioning and stable ablation electrode-endocardial contact. This procedure is currently guided by an analysis of intracardiac electrograms and fluoroscopy. However, the use of fluoroscopy does not allow the endocardium and certain anatomic landmarks to be identified and is associated with the hazards of radiation exposure.
Methods: Seventeen symptomatic patients were studied. A 10F 10-MHz intracardiac imaging catheter was used to visualize specific anatomic landmarks in the right atrium for directing the ablation electrode in 15 patients undergoing radiofrequency ablation of 19 arrhythmias and to assist with interatrial septal puncture in 3 patients.
Results: Continuous intracardiac imaging was performed for a mean +/- SD of 63.6 +/- 39.2 min and demonstrated distal electrode-endocardial tissue contact in 81 (60%) of 134 radiofrequency applications. Movement of the catheter was demonstrated during 36 (44%), microcavitations during 39 (48%) and thrombus during 15 (19%) of the 81 imaged applications. In 7 of 10 procedures for atrial flutter, successful ablation was directed at anatomic corridors in the right atrium visualized with intracardiac echocardiography. During ablation of atrial tachycardia, imaging identified abnormal atrial anatomy related to previous surgery and guided successful ablation of a reentrant tachycardia circulating around these anatomic obstacles. In two procedures for slow pathway modification of atrioventricular node reentrant tachycardia, intracardiac echocardiography confirmed catheter stability at the tricuspid annulus anterior to the coronary sinus.
Conclusions: During catheter ablation, intracardiac echocardiography augments fluoroscopy by visualizing anatomic landmarks, ensuring stable endocardial contact and assisting in transseptal puncture. Ablation of typical atrial flutter can be successfully directed at anatomic corridors identified using intracardiac imaging.
Similar articles
-
Radiofrequency catheter ablation guided by intracardiac echocardiography.Circulation. 1994 Mar;89(3):1301-5. doi: 10.1161/01.cir.89.3.1301. Circulation. 1994. PMID: 8124819
-
Curing reentrant atrial arrhythmias. Targeting protected zones of slow conduction by catheter ablation.J Electrocardiol. 1993;26 Suppl:194-203. J Electrocardiol. 1993. PMID: 8189124
-
Radiofrequency catheter ablation of atrial arrhythmias. Results and mechanisms.Circulation. 1994 Mar;89(3):1074-89. doi: 10.1161/01.cir.89.3.1074. Circulation. 1994. PMID: 8124793
-
The utility of intracardiac echocardiography in interventional electrophysiology.Curr Cardiol Rep. 2000 Jul;2(4):329-34. doi: 10.1007/s11886-000-0089-z. Curr Cardiol Rep. 2000. PMID: 10953267 Review.
-
Localization of the origin of the atrioventricular junctional rhythm induced during selective ablation of slow-pathway conduction in patients with atrioventricular node reentrant tachycardia.Am Heart J. 1996 May;131(5):937-46. doi: 10.1016/s0002-8703(96)90176-3. Am Heart J. 1996. PMID: 8615313 Review.
Cited by
-
Ablation of atrial fibrillation.Curr Cardiol Rep. 2002 Sep;4(5):379-87. doi: 10.1007/s11886-002-0037-1. Curr Cardiol Rep. 2002. PMID: 12169234 Review.
-
Intracardiac echocardiography: gross anatomy and magnetic resonance correlations and validations.Int J Cardiovasc Imaging. 2005 Aug;21(4):391-401. doi: 10.1007/s10554-004-5809-9. Int J Cardiovasc Imaging. 2005. PMID: 16047120
-
New applications of intracardiac echocardiography: assessment of coronary blood flow by colour and pulsed Doppler imaging in dogs.Heart. 2002 Sep;88(3):283-8. doi: 10.1136/heart.88.3.283. Heart. 2002. PMID: 12181224 Free PMC article.
-
Feedforward Coordinate Control of a Robotic Cell Injection Catheter.Cell Transplant. 2017 Aug;26(8):1319-1330. doi: 10.1177/0963689717720294. Cell Transplant. 2017. PMID: 28901195 Free PMC article.
-
Challenges and implementation of radiation-force imaging with an intracardiac ultrasound transducer.IEEE Trans Ultrason Ferroelectr Freq Control. 2007 May;54(5):996-1009. doi: 10.1109/tuffc.2007.345. IEEE Trans Ultrason Ferroelectr Freq Control. 2007. PMID: 17523564 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources