Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 1994:217:21-36.
doi: 10.3109/00365519409095208.

Characterization of a new ion selective electrode for ionized magnesium in whole blood, plasma, serum, and aqueous samples

Affiliations
Comparative Study

Characterization of a new ion selective electrode for ionized magnesium in whole blood, plasma, serum, and aqueous samples

B T Altura et al. Scand J Clin Lab Invest Suppl. 1994.

Abstract

Results from a novel ion selective electrode (ISE) for ionized magnesium (Mg2+) correlate well with atomic absorption spectroscopy on aqueous solutions containing from 0.1-3.0 mmol MgCl2/L. Day to day precision (coefficient of variation) of the electrode on protein-based controls is < 4%; aqueous-based controls < 6%. The new ISE is selective for Mg2+ with a selectivity constant for Ca2+ (KMgCa) of 8 x 10(-2). Adding pathophysiologic concentrations of Cd2+, Ca2+, Cu2+, Fe3+, K+, Na+, or Zn2+ to serum and aqueous solutions gave negligible to minimal changes in measured Mg2+. Ligand binding studies in aqueous solution indicate that pathophysiologic concentrations of different anions (e.g. heparin, lactate, bicarbonate, phosphate, acetate and sulfate) bind to Mg2+, effectively reducing its concentration in solution. Likewise, silicon (as either found in Vacuutainer tubes or as chlorosilane) failed to exert any significant effect on measured Mg2+. Addition of Intralipid (up to 500 mg/dL) gave negligible to minimal changes in Mg2+. Mg2+ measurements on whole blood, plasma, and serum for a given human subject's samples are virtually identical, at least within the reference range for Mg2+. Typically, Mg2+ is 71% of TMg, but varies from subject to subject; i.e. Mg2+ cannot be predicted from TMg. Clinical studies revealed that the Mg2+/TMg ratio could be remarkably consistent in sequential samples (e.g., throughout the course of coronary bypass surgery) taken from one patient, but that this ratio could differ dramatically from the ratio in sequential samples taken from another. Mg2+ is held within a narrow range (0.53-0.67 mmol/L) in normal, healthy subjects when compared to TMg (0.70-0.96 mmol/L).

PubMed Disclaimer

Publication types

LinkOut - more resources