. 1994 Nov;22(11):1805-8.

Noninvasive monitoring of end-tidal CO2 via nasal cannulas in spontaneously breathing children during the perioperative period

Affiliations
  • PMID: 7956285

Noninvasive monitoring of end-tidal CO2 via nasal cannulas in spontaneously breathing children during the perioperative period

J D Tobias et al. Crit Care Med. 1994 Nov.

Abstract

Objective: To determine the correlation between end-tidal CO2 and PaCO2 values measured via nasal cannulas in spontaneously breathing children during the perioperative period.

Design: Prospective evaluation.

Setting: Pediatric intensive/intermediate care unit in a tertiary care referral center.

Patients: Thirty postoperative surgical and trauma patients aged < or = 18 yrs (average age 7.8 yrs [range 6 months to 16 yrs] and average weight 28.3 kg (range 8.5 to 69).

Measurements and main results: Spontaneously breathing, nonintubated patients with an arterial cannula in place were selected for study. End-tidal CO2 was sampled from nasal cannulas by a sidestream aspirator and was estimated by infrared spectroscopy. The difference between PaCO2 and end-tidal CO2 was compared using linear regression analysis. A total of 55 blood gas measurements were obtained on the 30 patients. The PaCO2 to end-tidal CO2 gradient was < or = 4 torr in 54 of the 55 samples. The mean PaCO2 was 39.5 +/- 3.3 torr (5.27 +/- 0.44 kPa) with a mean end-tidal CO2 value of 39.7 +/- 3.8 torr (5.29 +/- 0.51 kPa). Linear regression analysis of arterial vs. end-tidal CO2 yielded a slope of 0.992 and p = .0001.

Conclusions: End-tidal CO2 measurement by infrared spectroscopy provided an accurate estimation of PaCO2 in this patient population. Its use may limit the need for invasive monitoring and/or repeated arterial blood gas analysis.