Accuracy of fingerstick glucose determination in patients receiving CPR
- PMID: 7973887
- DOI: 10.1097/00007611-199411000-00003
Accuracy of fingerstick glucose determination in patients receiving CPR
Abstract
Recent demonstration of the dangers of hyperglycemia in the setting of cerebral hypoperfusion has resulted in recommendations to abandon the long-standing practice of empiric administration of dextrose to patients with altered mental status. Patients receiving basic cardiopulmonary resuscitation (CPR) have been identified as being at particular risk from inappropriate administration of dextrose. We conducted a prospective trial to determine whether the blood glucose of 50 patients receiving CPR could be accurately assessed using a portable rapid-reagent device in the emergency department (ED). We analyzed both capillary and venous blood samples in the rapid-reagent system, using laboratory analysis of venous blood to define the patient's true blood glucose level. Capillary blood analysis identified 8 patients as hypoglycemic; only 3 of these patients were truly hypoglycemic, and 2 patients identified as hypoglycemic by fingerstick were actually hyperglycemic on laboratory analysis. The only instance in which a patient was classified differently by venous rapid-reagent and laboratory analysis occurred when a patient assessed as normoglycemic (169 mg/dL) by the laboratory had a venous rapid-reagent glucose reading of 182 mg/dL (hyperglycemic). We recommend that venous (rather than capillary) blood be used to analyze blood glucose in all patients receiving CPR before dextrose administration is considered.
Comment in
-
Fingerstick glucose determination and CPR.South Med J. 1995 Jun;88(6):697-8. doi: 10.1097/00007611-199506000-00026. South Med J. 1995. PMID: 7777898 No abstract available.
Similar articles
-
Fingerstick glucose determination and CPR.South Med J. 1995 Jun;88(6):697-8. doi: 10.1097/00007611-199506000-00026. South Med J. 1995. PMID: 7777898 No abstract available.
-
Fingerstick glucose determination in shock.Ann Intern Med. 1991 Jun 15;114(12):1020-4. doi: 10.7326/0003-4819-114-12-1020. Ann Intern Med. 1991. PMID: 2029097
-
Accuracy of reagent strips in detecting hypoglycemia in the emergency department.Ann Emerg Med. 1998 Sep;32(3 Pt 1):305-9. doi: 10.1016/s0196-0644(98)70005-8. Ann Emerg Med. 1998. PMID: 9737491
-
50% dextrose: antidote or toxin?Ann Emerg Med. 1990 Jun;19(6):683-7. doi: 10.1016/s0196-0644(05)82479-5. Ann Emerg Med. 1990. PMID: 2188538 Review.
-
Glucose.Emerg Med Clin North Am. 1986 Feb;4(1):193-205. Emerg Med Clin North Am. 1986. PMID: 3512245 Review.
Cited by
-
Comparing Self Monitoring Blood Glucose Devices and Laboratory Tests: Over 25 Years Experience.Cureus. 2019 Dec 1;11(12):e6268. doi: 10.7759/cureus.6268. Cureus. 2019. PMID: 31903305 Free PMC article.
-
Suitability of capillary blood glucose analysis in patients receiving vasopressors.Am J Crit Care. 2013 Sep;22(5):423-9. doi: 10.4037/ajcc2013692. Am J Crit Care. 2013. PMID: 23996422 Free PMC article.
-
A review of hypoglycemia and dextrose treatment in patients with cardiac arrest.Clin Exp Emerg Med. 2025 Mar;12(1):90-93. doi: 10.15441/ceem.24.305. Epub 2025 Jan 14. Clin Exp Emerg Med. 2025. PMID: 39806987 Free PMC article. No abstract available.
-
Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients.Intensive Care Med. 2005 Jan;31(1):142-5. doi: 10.1007/s00134-004-2500-5. Epub 2004 Nov 23. Intensive Care Med. 2005. PMID: 15565362 Clinical Trial.
-
The administration of dextrose during in-hospital cardiac arrest is associated with increased mortality and neurologic morbidity.Crit Care. 2015 Apr 10;19(1):160. doi: 10.1186/s13054-015-0867-z. Crit Care. 2015. PMID: 25887120 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical