Mechanical ventilators optimized for pediatric use decrease work of breathing and oxygen consumption during pressure-support ventilation
- PMID: 7988130
Mechanical ventilators optimized for pediatric use decrease work of breathing and oxygen consumption during pressure-support ventilation
Abstract
Objectives: a) To investigate whether the patient work of breathing needed to trigger inspiration is affected by the type of ventilator delivering pressure-support ventilation for mechanically ventilated pediatric patients. b) To determine whether changes in oxygen consumption (VO2) trend with changes in work of breathing and would thus be helpful in tracking work of breathing.
Design: Prospective study.
Setting: Pediatric intensive care unit at a university hospital.
Patients: Nine mechanically ventilated patients (2 to 75 months of age).
Interventions: While maintaining a constant pressure-support ventilation level, patients were alternately supported with the Siemens Servo 900C, the Bird VIP, and the Newport Wave E200 ventilators in random order.
Measurements and main results: Work of breathing, defined as the integral of the pressure-volume curve corresponding to negative pressure, was calculated with a pulmonary monitoring system. VO2 was measured with a metabolic cart. Patient distress levels were assessed using the COMFORT scale, a behavioral scoring system. Mean values (20 breaths/patient) for measured variables with each ventilator were compared using analysis of variance and Scheffé tests, with p < .05 indicating statistical significance. The lowest VO2 (103 +/- 35 mL/min/m2) and work of breathing (24 +/- 15 g.cm/m2) were achieved with the Bird VIP ventilator and were significantly (p < .05) lower than those values obtained with either the Siemens Servo 900C (VO2 147 +/- 33 mL/min/m2; work of breathing 49 +/- 18 g.cm/m2) or the Newport Wave E200 (VO2 122 +/- 33 mL/min/m2; work of breathing 35 +/- 15 g.cm/m2). Also, the values of work of breathing and VO2 obtained using the Newport Wave E200 were significantly (p < .05) lower than those values obtained using the Servo 900C. No change in behavioral distress occurred when the ventilators were changed. In all patients, there was a clear similarity in the trends of VO2 and work of breathing.
Conclusions: We conclude that VO2 and work of breathing may be reduced significantly using the latest generation of mechanical ventilators optimized for infant and pediatric use. Because work of breathing is less with the Bird VIP than the other two ventilators tested, leading to a corresponding decrease in VO2, we suggest that the Bird VIP better adapts the patient to the ventilator and may facilitate weaning from ventilatory support. We also suggest that changes in VO2 might be helpful in tracking changes in work of breathing.
Comment in
-
Mechanical ventilators optimized for pediatric use decrease work of breathing and oxygen consumption during pressure-support ventilation.Crit Care Med. 1995 Nov;23(11):1931-3. doi: 10.1097/00003246-199511000-00025. Crit Care Med. 1995. PMID: 7587274 No abstract available.
Similar articles
-
Comparison of total resistive work of breathing in two generations of ventilators in an animal model.Pediatr Pulmonol. 1996 Jul;22(1):58-66. doi: 10.1002/(SICI)1099-0496(199607)22:1<58::AID-PPUL8>3.0.CO;2-G. Pediatr Pulmonol. 1996. PMID: 8856804
-
Triggering delay time and work of breathing in three paediatric patient-triggered ventilators.Can J Anaesth. 1998 Mar;45(3):261-5. doi: 10.1007/BF03012913. Can J Anaesth. 1998. PMID: 9579266 Clinical Trial.
-
Pressure support and pressure assist/control: are there differences? An evaluation of the newest intensive care unit ventilators.Respir Care. 2000 Oct;45(10):1169-81. Respir Care. 2000. PMID: 11054896
-
Ventilator waveform interpretation in mechanically ventilated small animals.J Vet Emerg Crit Care (San Antonio). 2011 Oct;21(5):496-514. doi: 10.1111/j.1476-4431.2011.00673.x. J Vet Emerg Crit Care (San Antonio). 2011. PMID: 22316197 Review.
-
Strategies to minimize breathing effort during mechanical ventilation.Crit Care Clin. 1990 Jul;6(3):635-61. Crit Care Clin. 1990. PMID: 2198999 Review.
Cited by
-
Advances in neonatal conventional ventilation.Arch Dis Child Fetal Neonatal Ed. 1996 Sep;75(2):F135-40. doi: 10.1136/fn.75.2.f135. Arch Dis Child Fetal Neonatal Ed. 1996. PMID: 8949700 Free PMC article. Review. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources