Keratoconus and contact lens-induced corneal warpage analysis using the keratomorphic diagram
- PMID: 8002240
Keratoconus and contact lens-induced corneal warpage analysis using the keratomorphic diagram
Abstract
Purpose: Videokeratography of early keratoconus may be difficult to distinguish from contact lens-induced corneal warpage, even by experienced examiners. Furthermore, topographic irregularity may be judged inconsistently if quantitative standards are not applied. Quantitative measures based on videokeratographic data were developed and evaluated to determine if improved corneal topographic classification can be achieved.
Methods: The Corneal Irregularity Coefficient (CIC) and Corneal Power Coefficient (CPC) were derived from multiple measures of mean corneal power and its variance for 207 videokeratographs of normal, warped, keratoconus, and keratoconus-suspect corneas. CIC was plotted against CPC, creating a distribution of points representing all maps that tended to be grouped according to surface conditions (the Keratomorphic Diagram). Normal, steep, abnormal, and warped zones were defined by CIC and CPC cutoff values chosen to distinguish normal from keratoconus corneas graphically.
Results: Seventy of 76 normal corneas were grouped in the normal zone and 6 in the steep zone; 84 of 84 keratoconus corneas were grouped in the abnormal zone; 35 of 35 contact lens-induced warpage cases were grouped in the warped zone; and 10 of 12 keratoconus-suspect corneas were grouped in the warped zone, with 2 in the abnormal zone. Serially plotted data of keratoconus progression and warpage regression demonstrated that the vector displacement of CIC and CPC values may provide a potentially useful means of distinguishing contact lens-induced warpage from keratoconus-suspect corneas.
Conclusion: The Keratomorphic Diagram aids in classifying and comparing corneal shape by plotting indices along axes with easily recalled scales. The diagram may become a useful tool to assess presurgical corneal surface instability and postoperative progression of corneal shape change due to healing.
Similar articles
-
Goodness-of-prediction of Zernike polynomial fitting to corneal surfaces.J Cataract Refract Surg. 2005 Dec;31(12):2350-5. doi: 10.1016/j.jcrs.2005.05.025. J Cataract Refract Surg. 2005. PMID: 16473230
-
Automated keratoconus screening with corneal topography analysis.Invest Ophthalmol Vis Sci. 1994 May;35(6):2749-57. Invest Ophthalmol Vis Sci. 1994. PMID: 8188468
-
Differentiating contact lens induced warpage from true keratoconus using corneal topography.CLAO J. 1999 Apr;25(2):114-22. CLAO J. 1999. PMID: 10344298
-
Quantitative analysis of corneal microstructure in keratoconus utilising in vivo confocal microscopy.Eye (Lond). 2007 May;21(5):614-23. doi: 10.1038/sj.eye.6702286. Epub 2006 Feb 24. Eye (Lond). 2007. PMID: 16498438 Review.
-
Contact lens fitting in keratoconus.Compr Ophthalmol Update. 2006 Mar-Apr;7(2):47-52. Compr Ophthalmol Update. 2006. PMID: 16709339 Review.
Cited by
-
Five-year corneal cross-linking outcomes: A Save Sight Keratoconus Registry Study.Clin Exp Ophthalmol. 2023 Jan;51(1):9-18. doi: 10.1111/ceo.14177. Epub 2022 Oct 24. Clin Exp Ophthalmol. 2023. PMID: 36240047 Free PMC article.
-
Longitudinal changes in corneal irregular astigmatism and visual acuity in eyes with keratoconus.Jpn J Ophthalmol. 2007 Jul-Aug;51(4):265-9. doi: 10.1007/s10384-007-0453-2. Epub 2007 Aug 3. Jpn J Ophthalmol. 2007. PMID: 17660986
-
Proposing an ensemble learning model based on neural network and fuzzy system for keratoconus diagnosis based on Pentacam measurements.Int Ophthalmol. 2021 Dec;41(12):3935-3948. doi: 10.1007/s10792-021-01963-2. Epub 2021 Jul 28. Int Ophthalmol. 2021. PMID: 34322847
-
Differentiating Between Contact Lens Warpage and Keratoconus Using OCT Maps of Corneal Mean Curvature and Epithelial Thickness.J Refract Surg. 2022 Feb;38(2):112-119. doi: 10.3928/1081597X-20211116-01. Epub 2022 Feb 1. J Refract Surg. 2022. PMID: 35156455 Free PMC article.
-
Royston-Parmar flexible parametric survival model to predict the probability of keratoconus progression to corneal transplantation.Eye (Lond). 2020 Apr;34(4):657-662. doi: 10.1038/s41433-019-0554-4. Epub 2019 Aug 28. Eye (Lond). 2020. PMID: 31462761 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous