Enzymatic glucose sensors. Improved long-term performance in vitro and in vivo
- PMID: 8003752
Enzymatic glucose sensors. Improved long-term performance in vitro and in vivo
Abstract
We studied the long-term in vitro and in vivo performance of enzyme electrode glucose sensors. Single commercially produced enzyme-active membranes remained functional for estimating glucose in vitro for 14-36 months. These membranes were implanted subcutaneously in rats for 1 year and, upon explanation, remained functional for measuring glucose in vitro. Sensors with these membranes plus an additional outer membrane with lower glucose permeability allowed glucose monitoring in the low oxygen tension of subcutaneous tissue. These sensors were surgically implanted in three nondiabetic dogs. Each sensor implant was coupled to a radio transmitter to allow continuous long-term glucose monitoring in these awake unrestrained dogs. In vivo sensor performance was evaluated by intravenous glucose infusion, with reference blood glucose determinations made in the clinical laboratory. These subcutaneously implanted sensors tracked changes in plasma glucose for up to 12 weeks. The in vivo initial response for three sensor implants was approximately 35 sec (n = 8). Sensor peak response to glucose after bolus infusion ranged from 3 to 14 min. Stability of sensor sensitivity within +/- 15% for more than 1 month was demonstrated in two of the dogs. Sensor lifetime was limited not by loss of enzyme activity, but by biodegradation of the outermost polyurethane membrane. The findings suggest that long-term continuous monitoring of blood glucose using a subcutaneously implanted enzyme electrode sensor may be possible.
Similar articles
-
[Subcutaneously implantable glucose sensors in patients with diabetes mellitus; still many problems].Ned Tijdschr Geneeskd. 2002 Jul 13;146(28):1313-6. Ned Tijdschr Geneeskd. 2002. PMID: 12148218 Review. Dutch.
-
Continuous glucose monitoring in interstitial fluid using glucose oxidase-based sensor compared to established blood glucose measurement in rats.Anal Chim Acta. 2007 Jan 2;581(1):7-12. doi: 10.1016/j.aca.2006.08.016. Epub 2006 Aug 18. Anal Chim Acta. 2007. PMID: 17386418
-
A subcutaneous glucose sensor with improved longevity, dynamic range, and stability of calibration.Diabetes Care. 2000 Feb;23(2):208-14. doi: 10.2337/diacare.23.2.208. Diabetes Care. 2000. PMID: 10868833
-
The GOD-H2O2-electrode as an approach to implantable glucose sensors.Horm Metab Res Suppl. 1988;20:26-9. Horm Metab Res Suppl. 1988. PMID: 3248787
-
The glucose sensor: the missing link in diabetes therapy.Horm Metab Res Suppl. 1990;24:154-64. Horm Metab Res Suppl. 1990. PMID: 2272621 Review.
Cited by
-
Percutaneous window chamber method for chronic intravital microscopy of sensor-tissue interactions.J Diabetes Sci Technol. 2008 Nov;2(6):977-83. doi: 10.1177/193229680800200608. J Diabetes Sci Technol. 2008. PMID: 19885287 Free PMC article.
-
Modulation of fibroblast inflammatory response by surface modification of a perfluorinated ionomer.Biointerphases. 2011 Jun;6(2):43-53. doi: 10.1116/1.3583535. Biointerphases. 2011. PMID: 21721839 Free PMC article.
-
Technologies for continuous glucose monitoring: current problems and future promises.J Diabetes Sci Technol. 2010 Nov 1;4(6):1540-62. doi: 10.1177/193229681000400632. J Diabetes Sci Technol. 2010. PMID: 21129353 Free PMC article. Review.
-
Intravital microscopy evaluation of angiogenesis and its effects on glucose sensor performance.J Biomed Mater Res A. 2010 Jun 15;93(4):1348-57. doi: 10.1002/jbm.a.32630. J Biomed Mater Res A. 2010. PMID: 19911378 Free PMC article.
-
Emerging synergy between nanotechnology and implantable biosensors: a review.Biosens Bioelectron. 2010 Mar 15;25(7):1553-65. doi: 10.1016/j.bios.2009.12.001. Epub 2009 Dec 11. Biosens Bioelectron. 2010. PMID: 20042326 Free PMC article. Review.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical