Pediatric defibrillation: current flow is improved by using "adult" electrode paddles
- PMID: 8008545
Pediatric defibrillation: current flow is improved by using "adult" electrode paddles
Abstract
Objective: Current flow, the major determinant of defibrillation success, depends on delivered energy and transthoracic impedance. Previous experimental data suggest that transthoracic impedance is higher using "pediatric" electrode paddles compared to the larger "adult" electrode paddles. There are few data from actual shocks to support the experimental studies. The purpose of this study was to measure transthoracic impedance during actual shock delivery and to determine the optimal electrode paddle size for pediatric defibrillation and cardioversion.
Methods: We prospectively evaluated all shocks of < or = 20 joules given to pediatric patients at the University of Iowa from 1988 to 1992. Data collected included energy selected by the operator, energy delivered by the defibrillator, peak current flow and transthoracic impedance. Data were analyzed by unpaired t test and linear regression.
Results: Fifty-five shocks were delivered to 20 patients, age newborn to 8 years. Thirty-seven shocks were given with "pediatric" electrode paddles (surface area 21 cm2) and 18 with "adult" electrode paddles (surface area 83 cm2). Selected energy correlated well with delivered energy (r = .98, P < .0001). Delivered energy using "pediatric" electrodes did not differ from that delivered with "adult" electrodes (8.0 +/- 0.9 joules vs 10.5 +/- 1.2 joules P > .1). However transthoracic impedance was significantly higher with "pediatric" electrodes (78.1 +/- 4.4 ohms vs 54.6 +/- 2.7 ohms P < .0008), resulting in lower peak current flow through "pediatric" electrode paddles (6.2 +/- 0.5 amps vs 8.7 +/- 0.5 amps P < .002). There was no correlation between joules/kg and peak current flow (r = .26, P > .05).
Conclusion: Use of "pediatric" electrode paddles results in higher transthoracic impedance and thus lower peak current flow. In pediatric defibrillation, larger "adult" electrode paddles should be used as soon as chest size permits (approximately 10 kg). Lower transthoracic impedance results in higher current flow that facilitates cardioversion and defibrillation.
Similar articles
-
Pediatric defibrillation: importance of paddle size in determining transthoracic impedance.Pediatrics. 1988 Dec;82(6):914-8. Pediatrics. 1988. PMID: 3186384
-
Disposable defibrillator electrodes.Heart Lung. 1977 Jan-Feb;6(1):127-30. Heart Lung. 1977. PMID: 583893
-
Transthoracic impedance study with large self-adhesive electrodes in two conventional positions for defibrillation.Physiol Meas. 2006 Oct;27(10):1009-22. doi: 10.1088/0967-3334/27/10/007. Epub 2006 Aug 25. Physiol Meas. 2006. PMID: 16951460
-
[The implantable automatic cardioverter-defibrillator].Herz. 1990 Apr;15(2):111-25. Herz. 1990. PMID: 2188891 Review. German.
-
[Electrical defibrillation and cardioversion].Medicina (Kaunas). 2005;41(10):892-9. Medicina (Kaunas). 2005. PMID: 16272838 Review. Lithuanian.
Cited by
-
Ventricular fibrillation and defibrillation.Arch Dis Child. 2007 Oct;92(10):916-21. doi: 10.1136/adc.2006.115618. Arch Dis Child. 2007. PMID: 17895341 Free PMC article. Review.
-
A Systematic Review of the Transthoracic Impedance during Cardiac Defibrillation.Sensors (Basel). 2022 Apr 6;22(7):2808. doi: 10.3390/s22072808. Sensors (Basel). 2022. PMID: 35408422 Free PMC article.
-
Pediatric basic and advanced life support: 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science with Treatment Recommendations.Pediatrics. 2010 Nov;126(5):e1261-318. doi: 10.1542/peds.2010-2972A. Epub 2010 Oct 18. Pediatrics. 2010. PMID: 20956433 Free PMC article. No abstract available.
-
Defibrillation in children.J Emerg Trauma Shock. 2010 Jul;3(3):261-6. doi: 10.4103/0974-2700.66526. J Emerg Trauma Shock. 2010. PMID: 20930970 Free PMC article.
-
Analysis of transthoracic impedance during real cardiac arrest defibrillation attempts in older children and adolescents: are stacked-shocks appropriate?Resuscitation. 2010 Nov;81(11):1540-3. doi: 10.1016/j.resuscitation.2010.07.011. Epub 2010 Aug 13. Resuscitation. 2010. PMID: 20708836 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources