Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS polyethylene liner
- PMID: 8014641
- DOI: 10.1016/0883-5403(94)90059-0
Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS polyethylene liner
Abstract
A retrospective review was performed of 94 consecutive Anatomic Medullary Locking Plus (AML+, DePuy, Warsaw, IN) cementless acetabular components implanted between January 1988 and January 1990. All acetabular cups utilized the Acetabular Cup System (ACS) polyethylene liner (Depuy). Of these hips, 72 had been followed for more than 2 years (average, 43 months). There have been 15 (21%) clinical failures, all due to catastrophic wear of the acetabular polyethylene component surfaces. Of the 15 failures, clinical symptoms were absent in 7, an audible squeak was present in 3, and 5 patients reported having pain. All clinical failures had accelerated acetabular wear necessitating revision. Patients in the failure population were younger (56 years vs 62 years) than the remainder of the patients, and had greater cup abduction angles (55 degrees vs 49 degrees). All of the failed ACS polyethylene components had a 32 mm inner diameter articulating surface. The AML + acetabular component outer diameter averaged 56 mm (range, 50-64 mm). Six of 15 failures occurred in cups 58 mm or larger. Acetabular wear in the failure group was 0.77 mm/y (average). The incidence of acetabular (78%) and femoral (71%) osteolysis in the failure population is alarming. The rate of failure (21%) at 46 months of the AML + acetabular component with the ACS polyethylene liner appears to be related to design. The ACS design is flawed by a lack of hemispherical geometry, leading to failure at the superior rim with penetration of the femoral head through the polyethylene and against the metal shell.(ABSTRACT TRUNCATED AT 250 WORDS)
Comment in
-
Catastrophic polyethylene wear in hip arthroplasty.J Arthroplasty. 1994 Apr;9(2):117-8. doi: 10.1016/0883-5403(94)90058-2. J Arthroplasty. 1994. PMID: 8014640 No abstract available.
-
Acetabular wear.J Arthroplasty. 1996 Jan;11(1):122. doi: 10.1016/s0883-5403(96)80173-2. J Arthroplasty. 1996. PMID: 8676113 No abstract available.
Similar articles
-
Long-term results of uncemented acetabular cups with an ACS polyethylene liner. a 14-16-year follow-up study.Int Orthop. 2007 Apr;31(2):205-10. doi: 10.1007/s00264-006-0151-6. Epub 2006 May 31. Int Orthop. 2007. PMID: 16736147 Free PMC article.
-
Severe polyethylene wear in uncemented acetabular cup system components: a report of 5 cases.J Arthroplasty. 1999 Aug;14(5):635-6. doi: 10.1016/s0883-5403(99)90090-6. J Arthroplasty. 1999. PMID: 10475567
-
Complete acetabular cup revision versus isolated liner exchange for polyethylene wear and osteolysis without loosening in cementless total hip arthroplasty.Arch Orthop Trauma Surg. 2011 Nov;131(11):1591-600. doi: 10.1007/s00402-011-1338-x. Epub 2011 Jun 19. Arch Orthop Trauma Surg. 2011. PMID: 21687959
-
Late failure of the polyethylene liner fixation in an uncemented total hip arthroplasty.J Arthroplasty. 1999 Apr;14(3):391-6. doi: 10.1016/s0883-5403(99)90070-0. J Arthroplasty. 1999. PMID: 10220198 Review.
-
[Reliability of porous coating metal-backed cups: advantages and adverse effect of a posterior elevated rim polyethylene liner].Rev Chir Orthop Reparatrice Appar Mot. 2005 Sep;91(5):432-8. doi: 10.1016/s0035-1040(05)84360-0. Rev Chir Orthop Reparatrice Appar Mot. 2005. PMID: 16351000 Review. French.
Cited by
-
Clinical biomechanics of wear in total hip arthroplasty.Iowa Orthop J. 2003;23:1-12. Iowa Orthop J. 2003. PMID: 14575243 Free PMC article.
-
Reduced wear in vitamin E-infused highly cross-linked polyethylene cups: 5-year results of a randomized controlled trial.Acta Orthop. 2021 Apr;92(2):151-155. doi: 10.1080/17453674.2020.1852785. Epub 2020 Dec 2. Acta Orthop. 2021. PMID: 33263447 Free PMC article. Clinical Trial.
-
Clinical and radiographic results and wear performance in different generations of a cementless porous-coated acetabular cup.Int Orthop. 2008 Apr;32(2):181-7. doi: 10.1007/s00264-006-0305-6. Epub 2007 Jan 31. Int Orthop. 2008. PMID: 17265157 Free PMC article.
-
Squeaking: Current knowledge and how to avoid it.Curr Rev Musculoskelet Med. 2013 Dec;6(4):342-9. doi: 10.1007/s12178-013-9181-z. Curr Rev Musculoskelet Med. 2013. PMID: 23925448 Free PMC article.
-
Long-term results of uncemented acetabular cups with an ACS polyethylene liner. a 14-16-year follow-up study.Int Orthop. 2007 Apr;31(2):205-10. doi: 10.1007/s00264-006-0151-6. Epub 2006 May 31. Int Orthop. 2007. PMID: 16736147 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources