A quantitative in vitro assessment of fit and screw fixation on the stability of a cementless hemispherical acetabular component
- PMID: 8014647
- DOI: 10.1016/0883-5403(94)90065-5
A quantitative in vitro assessment of fit and screw fixation on the stability of a cementless hemispherical acetabular component
Abstract
This investigation quantifies in vitro the effect of component fit, as well as the effect of adjuvant screw fixation, on the initial stability of cementless hemispherical titanium acetabular total hip arthroplasty components and assesses apposition of the acetabular components to bone. Six, fresh human hemipelvi (3 matched pairs) were harvested at autopsy. Titanium alloy acetabular components with a porous surface of commercially pure titanium fiber mesh (Harris Galante Porous acetabular components, Zimmer, Warsaw, IN) were used for implantation. Initially, each acetabulum was underreamed to achieve a 2 mm press-fit with the acetabular component. Pressure-sensitive film had been placed along the dome and medial wall at the bone-implant interface to assess the completeness of seating. After the implant was impacted into the acetabular cavity, relative motion between the implant and bone was measured during simulated single leg stance. Adjuvant fixation of the implant was then obtained with the insertion of four 6.5 mm cancellous screws. High-contrast roentgenograms of the specimens in multiple views were obtained after initial cup insertion and again after screw insertion. The stability of each implant under load was measured with four, three, two, one, and no screws in place. Further reaming of the bone was done to create a 1 mm press-fit. The sequence was then repeated. Further reaming was done to create an exact-fit and the sequence was repeated again. Under these conditions, 1 mm press-fit with or without screws provided the optimum combination of fit stability.(ABSTRACT TRUNCATED AT 250 WORDS)
Similar articles
-
The Harris-Galante porous acetabular component press-fit without screw fixation. Five-year radiographic analysis of primary cases.J Arthroplasty. 1994 Jun;9(3):235-42. doi: 10.1016/0883-5403(94)90077-9. J Arthroplasty. 1994. PMID: 8077971
-
Acetabular micromotion as a measure of initial implant stability in primary hip arthroplasty. An in vitro comparison of different methods of initial acetabular component fixation.J Arthroplasty. 1992 Dec;7(4):537-47. doi: 10.1016/s0883-5403(06)80076-8. J Arthroplasty. 1992. PMID: 1479374
-
Initial stability of cementless acetabular cups: press-fit and screw fixation interaction--an in vitro biomechanical study.Eur J Orthop Surg Traumatol. 2015 Apr;25(3):497-502. doi: 10.1007/s00590-014-1571-4. Epub 2014 Nov 25. Eur J Orthop Surg Traumatol. 2015. PMID: 25421640 Free PMC article.
-
Are additional screws required for press-fit fixation of cementless acetabular cups? A systematic review and meta-analysis.J Orthop Traumatol. 2022 Feb 10;23(1):9. doi: 10.1186/s10195-022-00629-8. J Orthop Traumatol. 2022. PMID: 35142933 Free PMC article.
-
The jumbo acetabular component for acetabular revision: Curtain Calls and Caveats.Bone Joint J. 2016 Jan;98-B(1 Suppl A):64-7. doi: 10.1302/0301-620X.98B1.36139. Bone Joint J. 2016. PMID: 26733644 Review.
Cited by
-
Cementless jumbo cups for revision of failed Furlong prosthesis. A case series.J Clin Orthop Trauma. 2020 Jan-Feb;11(1):56-61. doi: 10.1016/j.jcot.2018.08.009. Epub 2018 Aug 11. J Clin Orthop Trauma. 2020. PMID: 32001985 Free PMC article.
-
Modeling the debonding process of osseointegrated implants due to coupled adhesion and friction.Biomech Model Mechanobiol. 2023 Feb;22(1):133-158. doi: 10.1007/s10237-022-01637-7. Epub 2022 Oct 25. Biomech Model Mechanobiol. 2023. PMID: 36284076 Free PMC article.
-
Protrusio After Medial Acetabular Wall Breach in Total Hip Arthroplasty.Iowa Orthop J. 2015;35:99-107. Iowa Orthop J. 2015. PMID: 26361450 Free PMC article.
-
Cup press fit in uncemented THA depends on sex, acetabular shape, and surgical technique.Clin Orthop Relat Res. 2012 Nov;470(11):3014-23. doi: 10.1007/s11999-012-2381-9. Clin Orthop Relat Res. 2012. PMID: 22576930 Free PMC article.
-
Ex Vivo Evaluation of Cementless Acetabular Cup Stability Using Impact Analyses with a Hammer Instrumented with Strain Sensors.Sensors (Basel). 2017 Dec 27;18(1):62. doi: 10.3390/s18010062. Sensors (Basel). 2017. PMID: 29280982 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources