Accuracy of infrared ear thermometry and traditional temperature methods in young children
- PMID: 8039988
Accuracy of infrared ear thermometry and traditional temperature methods in young children
Abstract
Objective: To compare the accuracy of ear-based, rectal, and axillary temperature measurements in comparison to bladder temperature as a core reference.
Design: Repeated-measures comparison study.
Settings: Pediatric critical care settings in two tertiary care hospitals.
Patients: Thirty children, 1 to 45 months old (mean 16.6 months), who required bladder catheters for their care.
Outcome measures: Correlation and agreement (mean offset +/- SD) of ear-based, rectal, and axillary temperature measurements with bladder temperature.
Procedure: Ear-based measurements were made with three infrared thermometers in the core mode, both with and without an ear tug. All six readings were made in the same ear in randomized order. Bladder, rectal, and axillary temperatures were read from continuous digital displays immediately after each ear-based measurement.
Results: Ear-based readings correlated relatively well with bladder temperature (r = 0.80 to 0.87), but were lower by means of -0.3 degrees to -0.7 degrees C with moderately high variation (SD = 0.4 degrees to 0.5 degrees C) between children. Use of an ear tug did not affect the readings. Rectal temperature correlated well with bladder values (r = 0.93 to 0.97) and was usually slightly higher (mean offset = 0.2 +/- 0.2 [SD] degrees C), while axillary temperature correlated rather poorly (r = 0.59 to 0.64), with much lower and more variable readings (mean offset = 0.9 degrees +/- 0.6 degrees C). In regard to sensitivity, specificity, and predictive value in screening for fever, rectal readings performed very well, ear-based readings moderately well with some variation, and axillary readings poorly.
Conclusions: The findings suggest that the additive core-mode adjustments in infrared ear thermometers are too low for young children, an ear tug is not an essential part of measurement technique, rectal temperature closely reflects bladder temperature, and axillary temperature is low and highly variable.
Comment in
-
Accuracy of infrared ear thermometry and traditional temperature methods in young children.Heart Lung. 1995 May-Jun;24(3):260. doi: 10.1016/s0147-9563(05)80047-0. Heart Lung. 1995. PMID: 7622402 No abstract available.
Similar articles
-
Accuracy of infrared ear thermometry and other temperature methods in adults.Am J Crit Care. 1994 Jan;3(1):40-54. Am J Crit Care. 1994. PMID: 8118492
-
Reliability of infrared ear thermometry in the prediction of rectal temperature in older inpatients.J Clin Nurs. 2009 Feb;18(3):451-6. doi: 10.1111/j.1365-2702.2008.02565.x. J Clin Nurs. 2009. PMID: 19191993
-
Accuracy of tympanic temperature readings in children under 6 years of age.Pediatr Nurs. 1999 Jan-Feb;25(1):39-42. Pediatr Nurs. 1999. PMID: 10335248
-
Determination of oral temperature accuracy in adult critical care patients who are orally intubated.Heart Lung. 1994 Jul-Aug;23(4):300-7. Heart Lung. 1994. PMID: 7960855 Review.
-
The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis.Intern Emerg Med. 2021 Jun;16(4):1071-1083. doi: 10.1007/s11739-020-02556-0. Epub 2020 Nov 25. Intern Emerg Med. 2021. PMID: 33237494 Free PMC article.
Cited by
-
Temperature measurement in paediatrics.Paediatr Child Health. 2000 Jul;5(5):273-84. doi: 10.1093/pch/5.5.273. Paediatr Child Health. 2000. PMID: 20177532 Free PMC article. No abstract available.
-
Thermometry in paediatric practice.Arch Dis Child. 2006 Apr;91(4):351-6. doi: 10.1136/adc.2005.088831. Arch Dis Child. 2006. PMID: 16551792 Free PMC article. Review.
-
Assessment of axillary temperature for the evaluation of normal body temperature of healthy young adults at rest in a thermoneutral environment.J Physiol Anthropol. 2017 Feb 22;36(1):18. doi: 10.1186/s40101-017-0133-y. J Physiol Anthropol. 2017. PMID: 28228153 Free PMC article.
-
Reexamining age, race, site, and thermometer type as variables affecting temperature measurement in adults - A comparison study.BMC Nurs. 2003 Jun 15;2(1):1. doi: 10.1186/1472-6955-2-1. BMC Nurs. 2003. PMID: 12807535 Free PMC article.
-
Smart Patch for Skin Temperature: Preliminary Study to Evaluate Psychometrics and Feasibility.Sensors (Basel). 2021 Mar 6;21(5):1855. doi: 10.3390/s21051855. Sensors (Basel). 2021. PMID: 33800920 Free PMC article.