Clinical Trial
. 1994;15(2):137-40.

[Continuous guidance for venous punctures using a new pulsed Doppler probe: efficiency, safety]

[Article in French]
Affiliations
  • PMID: 8047199
Clinical Trial

[Continuous guidance for venous punctures using a new pulsed Doppler probe: efficiency, safety]

[Article in French]
B Branger et al. Nephrologie. 1994.

Abstract

Serious complications may occur after "blind" profound venous puncture in intensive care units. To secure these punctures, we designed a new fingertip pulsed Doppler (FPD) 5 MHz probe with a lateral to center indentation to guide the needle into the ultrasonic flux. A specific ultrasound analyzer indicating depth and diameter of the vessel was made for this use. The material was first tested in an experimental animal study. Results of animal venous punctures were successful 20/21. This material was then tested on patients with previous failure of "blind" punctures: results of 4 FPD punctures were successful in all 12 cases but one (catheterization not completed). We started a prospective multi-unit randomized study with various operators (junior residents, senior staff members), compared the success rate and type, the procedure duration of blind standard versus FPD punctures. We conclude in the safety and easy use of the FPD for central venous punctures.