Comparative Study
. 1994 May;83(5):366-72.

[Complications in transvenous and transvenous-subcutaneous implantable defibrillator systems]

[Article in German]
Affiliations
  • PMID: 8053246
Comparative Study

[Complications in transvenous and transvenous-subcutaneous implantable defibrillator systems]

[Article in German]
A Podczeck et al. Z Kardiol. 1994 May.

Abstract

Due to technical improvement using the transvenous-subcutaneous approach for lead fixation for cardioverter-defibrillator implantation, the incidence of device implantation has extended enormously. While a significant lower perioperative mortality in transvenously implanted systems compared to epicardially fixed implantable cardioverter-defibrillator (ICD) has been proven, perioperative as well as complications during follow-up are not analyzed systematically. In 59 patients, in whom transvenous-subcutaneous ICDs had been implanted, 3 patients showed bleeding complications in the subcutaneous patch area, 1 patient showed a bleeding in the device pocket, and in 1 patient a seroma in the subcutaneous patch region was observed necessitating surgical revisions. During follow-up of 10 +/- 7 months, lead dislocations were observed in 6 patients, in another 6 patients patch crinkling was observed. There was 1 patient with a lead fracture as well as 1 patient with a bleeding in the subcutaneous patch region. Despite the high efficacy of this therapy used in patients with life-threatening ventricular arrhythmias, these potential complications have to be considered.