Single X-ray absorptiometry of the forearm: precision, correlation, and reference data
- PMID: 8055369
- DOI: 10.1007/BF00301681
Single X-ray absorptiometry of the forearm: precision, correlation, and reference data
Abstract
The performance of a single X-ray absorptiometry (SXA) device incorporating an X-ray tube as a photon source was evaluated with respect to precision in vivo and in vitro, scan time, image quality, and correlation with an existing dual energy X-ray absorptiometry (DXA) device. SXA precision in vivo, expressed as a coefficient of variation (CV), was 0.66% for bone mineral content (BMC) and 1.05% for bone mineral density (BMD). Precision in vitro, based on 78 BMC measurements of a forearm phantom over 195 days, was 0.53%. Correlation with DXA at the 8 mm distal forearm site was high (r = 0.97 for BMC and r = 0.96 for BMD). A preliminary SXA reference database composed of 151 healthy Caucasian American women was developed to facilitate the interpretation of patient measurements. SXA scan time was 4 minutes and delivered a radiation exposure of 1.68 mrem. SXA image quality and spatial resolution were superior to SPA and comparable to DXA.
Similar articles
-
Single X-ray absorptiometry: performance characteristics and comparison with single photon absorptiometry.Osteoporos Int. 1995;5(5):377-81. doi: 10.1007/BF01622260. Osteoporos Int. 1995. PMID: 8800788
-
Bone densitometry of the forearm: comparison of single-photon and dual-energy X-ray absorptiometry.Bone. 1996 Jun;18(6):575-9. doi: 10.1016/8756-3282(96)00082-8. Bone. 1996. PMID: 8805999
-
Precision and accuracy of computed digital absorptiometry for assessment of bone density of the hand.Osteoporos Int. 1997;7(5):444-9. doi: 10.1007/s001980050031. Osteoporos Int. 1997. PMID: 9425502
-
Instrumental diagnosis of osteoporosis.Aging (Milano). 1998 Jun;10(3):240-8. doi: 10.1007/BF03339658. Aging (Milano). 1998. PMID: 9801734 Review.
-
[X-ray absorptiometry].Nihon Rinsho. 1998 Jun;56(6):1469-73. Nihon Rinsho. 1998. PMID: 9648467 Review. Japanese.
Cited by
-
Single X-ray absorptiometry: performance characteristics and comparison with single photon absorptiometry.Osteoporos Int. 1995;5(5):377-81. doi: 10.1007/BF01622260. Osteoporos Int. 1995. PMID: 8800788
-
Microscopic hematuria is associated with low bone mineral density in aged women and men.J Bone Miner Metab. 2009;27(2):251-4. doi: 10.1007/s00774-008-0027-5. Epub 2009 Jan 20. J Bone Miner Metab. 2009. PMID: 19153643
-
Comparison of bone densitometry of the phalanges, distal forearm and axial skeleton in early postmenopausal women participating in the EPIC Study.Osteoporos Int. 1996;6(4):308-13. doi: 10.1007/BF01623390. Osteoporos Int. 1996. PMID: 8883120 Clinical Trial.
-
Peripheral bone densitometry: an old friend revisited.Trans Am Clin Climatol Assoc. 1998;109:62-70; discussion 70-2. Trans Am Clin Climatol Assoc. 1998. PMID: 9601128 Free PMC article.
-
Differences in precision in bone mineral density measured by SXA and DXA: the NOREPOS study.Eur J Epidemiol. 2008;23(9):615-24. doi: 10.1007/s10654-008-9271-1. Epub 2008 Jul 2. Eur J Epidemiol. 2008. PMID: 18594994
References
MeSH terms
LinkOut - more resources
Medical