. 1994 Sep;192(3):637-43.
doi: 10.1148/radiology.192.3.8058927.

Radiation dosimetry for electron beam CT

Affiliations

Radiation dosimetry for electron beam CT

C H McCollough et al. Radiology. 1994 Sep.

Abstract

Purpose: To measure the radiation dose profile, multiple-scan average dose (MSAD), and computed tomography dose index (CTDI) for electron beam CT and to determine the accuracy of ionization-chamber and manufacturer estimates of patient dose.

Materials and methods: High-resolution dose profiles along the longitudinal axis were acquired at several positions within the scan plane with use of radiographic film. The full-width-at-half-maximum values, peak radiation dose, CTDI, and MSAD were calculated from the digitized film profiles. CTDI was also measured with an ionization chamber.

Results: The full-width-half-maximum value of the radiation profiles were significantly wider than the nominal scan width for the 6-mm single-section and 8-mm multisection modes. In the single-section mode, the CTDI underestimated the MSAD by 15%-30%. The multisection radiation profile was nonuniform and asymmetric.

Conclusion: Patient doses in electron beam CT are approximately 125% of the ionization-chamber CTDI measurements in the single-section mode. For the multisection mode, the average patient dose over the scan volume is approximately 70%-85% of the ionization-chamber CTDI measurements.

PubMed Disclaimer

Comment in

  • Coronary artery calcification.
    Doherty TM, Detrano RC. Doherty TM, et al. Radiology. 1995 May;195(2):576-7. doi: 10.1148/radiology.195.2.7598765. Radiology. 1995. PMID: 7598765 No abstract available.