. 1994 Sep;76(5):717-20.

Influence of metal implants on infection. An experimental study in rabbits

Affiliations
  • PMID: 8083257

Influence of metal implants on infection. An experimental study in rabbits

J Cordero et al. J Bone Joint Surg Br. 1994 Sep.

Abstract

We implanted cylinders of cobalt-chrome or titanium, with smooth or porous surfaces, into rabbit bones which had been inoculated with suspensions of Staphylococcus aureus in various doses. The bacterial concentration required to produce infection of porous-coated titanium implants was 2.5 times smaller than that necessary to infect implants with polished surfaces. Porous-coated cobalt-chromium implants required bacterial concentrations that were 40 times smaller than those needed to infect implants with polished surfaces, and 15 times smaller than those required to infect porous-coated titanium implants. The other advantages and disadvantages of the various implants, such as improved osseointegration, larger ion-release surfaces, surface wear and relative stiffness, must be weighed against the higher infection rates in the porous-coated implants, and particularly in the cobalt-chromium implants.

PubMed Disclaimer

Publication types

MeSH terms

LinkOut - more resources