Linac-accelerator-radiosurgery
- PMID: 8109312
- DOI: 10.1007/978-3-7091-9297-9_20
Linac-accelerator-radiosurgery
Abstract
A survey is given of the actual possibilities and limitations of the use of linear accelerators (Linac radiosurgery systems) for intra = cranial radiosurgery. Depending on the collimator size, spherical fields from 5-54 mm in diameter can be irradiated with dose gradients from 10% (large fields) to 20% (small fields) per millimeter distance between surface and treatment volume. This is comparable to the possibilities of Gamma-Knife and Proton-irradiation. Optimal mechanical adjustment of gantry and linac table are necessary for the required stability of the isocenter. Mechanical inaccuracy should be smaller than 0.8 mm. Advanced computerized 3D-treatment planning systems are indispensable prerequisites for accurate treatment and use of the flexibility of the linac system. Future developments are outlined.