Magnetic resonance imaging can cause focal heating in a nonuniform phantom

Affiliations

Magnetic resonance imaging can cause focal heating in a nonuniform phantom

P L Davis et al. IEEE Trans Biomed Eng. 1993 Dec.

Abstract

To test if the radiofrequency fields of a magnetic resonance imager could cause focal heating, two cylindrical phantoms were made from a mixture of agar and saline. The first phantom was uniform; the second was nonuniform in that a narrow bridge of agar was produced. Both phantoms were exposed to high levels of radiofrequency power (140 W) at 63 MHz and the temperature rises were measured. In the uniform phantom, the temperature increased as the radius increased. In the bridge phantom, the narrow bridge heated three times greater than at the opposite uniform periphery, and over five times the average of the uniform phantom. This experiment demonstrates that the radiofrequency fields of magnetic resonance imagers can cause focal heating if the exposed object is nonuniform. Since nonuniformity is present in the human body, as the radiofrequency power of magnetic resonance imaging techniques increases, focal heating in patients is a concern.

PubMed Disclaimer