Continuous, non-invasive volume-clamp blood pressure: determinants of performance
- PMID: 8133023
- DOI: 10.1097/00004872-199312000-00014
Continuous, non-invasive volume-clamp blood pressure: determinants of performance
Abstract
Objective: To test a prototype hydraulic, non-invasive, continuous finger blood pressure monitor based on the volume-clamp principle for procedure-related factors likely to influence precision.
Design: The influence of these factors was determined by repeatability of finger blood pressure measurement and the relationship to contralateral arm-cuff blood pressure.
Methods: Repeated blood pressure measurements from three different fingers were made in 60 subjects following re-initialization of the device and re-insertion of the finger into the cuff. Repeatability was assessed in relation to simultaneous arm-cuff readings. Drift in arm-finger discrepancy was measured over a 1-h period. Finger diameter, drug therapy and presence of peripheral vascular disease were correlated with arm-finger blood pressure difference.
Results: Repeatability coefficients (twice the SD of the arm-finger difference) across device re-initialization were large, but similar to parallel repeated arm blood pressure determinations: 17.6 and 17.1 mmHg for systolic blood pressure (SBP) and 13.9 and 13.6 mmHg for diastolic blood pressure (DBP), respectively. Withdrawing and re-inserting the finger reduced repeatability substantially, with a 50% increase in repeatability coefficient. A trend towards a progressive 9-mmHg increase was observed in overestimation of SBP over the 1-h period. Mean +/- SD pooled arm-finger blood pressure differences were -10.8 +/- 14.6 mmHg for SBP and 4.5 +/- 9.4 mmHg for DBP. Blood pressure measured in different fingers was similar on average, with repeatability no poorer than for re-insertion of the same finger. The presence of peripheral vascular disease in 15 subjects correlated with a smaller arm-finger difference for DBP.
Conclusions: Variations in positioning of the finger within the cuff influences blood pressure measurement during volume-clamp plethysmography, reducing its precision. Finger SBP exceeds brachial auscultatory readings and has similar precision.
Similar articles
-
Continuous noninvasive arterial pressure measurement using the volume clamp method: an evaluation of the CNAP device in intensive care unit patients.J Clin Monit Comput. 2015 Dec;29(6):807-13. doi: 10.1007/s10877-015-9670-2. Epub 2015 Mar 1. J Clin Monit Comput. 2015. PMID: 25726179
-
Noninvasive continuous blood pressure measurement from the finger: optimal measurement conditions and factors affecting reliability.J Clin Monit. 1987 Jan;3(1):6-13. doi: 10.1007/BF00770876. J Clin Monit. 1987. PMID: 3819798
-
Simultaneous validation of the SunTech 247 diagnostic station blood pressure measurement device according to the British Hypertension Society protocol, the International Protocol and the Association for the Advancement of Medical Instrumentation standards.Blood Press Monit. 2009 Oct;14(5):222-7. doi: 10.1097/MBP.0b013e328330c873. Blood Press Monit. 2009. PMID: 19741510
-
Fifteen years experience with finger arterial pressure monitoring: assessment of the technology.Cardiovasc Res. 1998 Jun;38(3):605-16. doi: 10.1016/s0008-6363(98)00067-4. Cardiovasc Res. 1998. PMID: 9747429 Review.
-
[Influence of the position of the arm on the level of clinical arterial pressure].Rev Med Interne. 1993 Feb;14(2):84-8. doi: 10.1016/s0248-8663(05)81256-1. Rev Med Interne. 1993. PMID: 8378630 Review. French.
Cited by
-
Asymmetric time-dependent model for the dynamic finger arterial pressure-volume relationship.Med Biol Eng Comput. 2006 Sep;44(9):829-34. doi: 10.1007/s11517-006-0090-9. Epub 2006 Aug 3. Med Biol Eng Comput. 2006. PMID: 16960748