Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern
- PMID: 8165071
Noninvasive monitoring of oxygenation in infants and children: practical considerations and areas of concern
Abstract
Objective: To review the technical principles of transcutaneous PO2 (tcPO2) monitors and pulse oximeters and to discuss methodological and practical issues related to the use of these devices in infants and children.
Design: Literature review and summary of personal experience.
Results: tcPO2 measurements are influenced by skin thickness, sensor temperature, amount of contact gel used, and state of peripheral perfusion. The average in vivo response time to a rapid decrease in PaO2 is approximately 16 seconds. Sensitivity to both hypoxemia (PaO2 < 50 mm Hg) and hyperoxemia (PaO2 > 80 to 100 mm Hg) is approximately 85%. Pulse oximeters require careful sensor placement (to avoid optical shunts) and adequate pulse pressures (> 20 mm Hg). They are prone to movement artifact. There are considerable differences in bias and precision between different brands, resulting, for example, in quite different upper alarm limits required to detect hyperoxemia reliably. The normal range of oxygen saturation measured by pulse oximetry, measured with one brand of pulse oximeter (Nellcor) during regular breathing, is 95% to 100% in preterm infants and 97% to 100% in full-term infants and children.
Conclusions: Because both devices have their specific shortfalls, they should ideally be used in combination, particularly in critically ill preterm neonates. Where this is not feasible or necessary, clinicians must be aware of the limitations of the device they are using.
Comment in
-
The newborn's SpO2: a routine vital sign whose time has come?Pediatrics. 1995 Jan;95(1):161-2. Pediatrics. 1995. PMID: 7770302 No abstract available.
Similar articles
-
Hyperoxemia in newborn infants: detection by pulse oximetry.Pediatrics. 1989 Aug;84(2):226-30. Pediatrics. 1989. PMID: 2748248
-
Pulse oximetry advantages in infants with bronchopulmonary dysplasia.Pediatrics. 1986 Nov;78(5):844-9. Pediatrics. 1986. PMID: 3763297
-
Pulse oximetry in very low birth weight infants with acute and chronic lung disease.Pediatrics. 1987 Apr;79(4):612-7. Pediatrics. 1987. PMID: 2434913
-
Noninvasive Monitoring and Assessment of Oxygenation in Infants.Clin Perinatol. 2019 Sep;46(3):417-433. doi: 10.1016/j.clp.2019.05.010. Epub 2019 Jun 10. Clin Perinatol. 2019. PMID: 31345538 Review.
-
Noninvasive monitoring of oxygen and carbon dioxide.Clin Chest Med. 1989 Jun;10(2):239-54. Clin Chest Med. 1989. PMID: 2661121 Review.
Cited by
-
Oxygen (Super) saturation?Indian J Anaesth. 2012 Nov;56(6):592-3. doi: 10.4103/0019-5049.104592. Indian J Anaesth. 2012. PMID: 23325954 Free PMC article. No abstract available.
-
Prophylactic nasal continuous positive airways pressure in newborns of 28-31 weeks gestation: multicentre randomised controlled clinical trial.Arch Dis Child Fetal Neonatal Ed. 2004 Sep;89(5):F394-8. doi: 10.1136/adc.2003.037010. Arch Dis Child Fetal Neonatal Ed. 2004. PMID: 15321956 Free PMC article. Clinical Trial.
-
Novel transcutaneous sensor combining optical tcPO2 and electrochemical tcPCO2 monitoring with reflectance pulse oximetry.Med Biol Eng Comput. 2020 Feb;58(2):239-247. doi: 10.1007/s11517-019-02067-x. Epub 2019 Nov 18. Med Biol Eng Comput. 2020. PMID: 31741291 Free PMC article.
-
Detection of hyperoxaemia in neonates: data from three new pulse oximeters.Arch Dis Child Fetal Neonatal Ed. 2002 Nov;87(3):F217-9. doi: 10.1136/fn.87.3.f217. Arch Dis Child Fetal Neonatal Ed. 2002. PMID: 12390996 Free PMC article.
-
Pulse oximetry, severe retinopathy, and outcome at one year in babies of less than 28 weeks gestation.Arch Dis Child Fetal Neonatal Ed. 2001 Mar;84(2):F106-10. doi: 10.1136/fn.84.2.f106. Arch Dis Child Fetal Neonatal Ed. 2001. PMID: 11207226 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources