Magnetic resonance imaging of coronary arteries: technique and preliminary results
- PMID: 8217439
- PMCID: PMC1025326
- DOI: 10.1136/hrt.70.4.315
Magnetic resonance imaging of coronary arteries: technique and preliminary results
Abstract
Background: Coronary artery imaging is an important investigation for the management of coronary artery disease. The only reliable technique presently available, x ray contrast angiography, is invasive and is associated with a small morbidity and mortality. Alternative non-invasive imaging would be useful, but the small calibre and tortuosity of the coronary vessels, and cardiac and respiratory motion create formidable imaging problems.
Objective: The development of rapid magnetic resonance imaging of the coronary arteries.
Patients: 21 healthy controls and five patients with coronary artery disease established by x ray contrast angiography, of whom two had undergone bypass grafting.
Methods: Magnetic resonance imaging was performed with gradient echoes and a segmented k-space technique, such that a complete image was acquired in 16 cardiac cycles during a breathhold. The signal from fat was suppressed and images were acquired in late diastole to reduce artefact from cardiac motion. An imaging strategy was developed for the proximal arteries, including longitudinal imaging from oblique planes defined according to the origins and the continuation of the arteries in the atrioventricular grooves or interventricular sulcus.
Results: Of the 26 subjects studied, 22 were imaged successfully. Identification of the artery was possible for the left main stem, left anterior descending, right coronary, and left circumflex arteries respectively in 95%, 91%, 95%, and 76%. The arterial diameter at the origin could be measured in 77%, 77%, 81%, and 63%. The mean (SD) arterial diameter in each case (4.8 (0.8), 3.7 (0.5), 3.9 (0.9), and 2.9 (0.6) mm) was not significantly different from reference values. The mean length of artery visualised was 10.4 (5.2), 46.7 (22.8), 53.7 (27.9), and 26.3 (17.5) mm. In 12 healthy men the total coronary area was 30.9 (9.2) mm2 and the ratio compared with body surface area was 16.4 (4.4) mm2m2 (both p = NS compared with reference values). In seven patients in whom x ray contrast coronary angiography was available, the proximal arterial diameter was 3.9 (1.1) mm measured by magnetic resonance and 3.7 (1.0) mm by x ray contrast angiography (p = NS). The mean difference between the measurements was 0.2 (0.5) mm, and the coefficient of variation was 13.7%. All five occluded coronary arteries were identified, as were all three vein grafts. In two patients insertion of the graft into the native arteries was identified.
Conclusions: Magnetic resonance coronary angiography is feasible. Good results were obtained by a breath-hold, fat suppression technique, gated to late diastole. Arterial occlusions and vein grafts were readily identified. Further studies are required to establish its value in the detection of coronary stenosis and to develop the measurement of coronary flow velocity which could be used to quantify the severity of the stenosis.
Similar articles
-
Breath-hold MR cine angiography of coronary arteries in healthy volunteers: value of multiangle oblique imaging planes.AJR Am J Roentgenol. 1994 Sep;163(3):533-7. doi: 10.2214/ajr.163.3.8079838. AJR Am J Roentgenol. 1994. PMID: 8079838
-
Assessment of coronary artery stenosis by magnetic resonance imaging.Heart. 1996 Feb;75(2):127-33. doi: 10.1136/hrt.75.2.127. Heart. 1996. PMID: 8673749 Free PMC article.
-
Fat-suppressed breath-hold magnetic resonance coronary angiography.Circulation. 1993 Jan;87(1):94-104. doi: 10.1161/01.cir.87.1.94. Circulation. 1993. PMID: 8419029
-
Coronary imaging using MRI.Herz. 1996 Apr;21(2):97-105. Herz. 1996. PMID: 8682443 Review.
-
Magnetic resonance coronary angiography in heart transplant recipients.Coron Artery Dis. 1996 Aug;7(8):591-7. doi: 10.1097/00019501-199608000-00006. Coron Artery Dis. 1996. PMID: 8922887 Review.
Cited by
-
Ischemic heart disease: value of MR techniques.Int J Card Imaging. 1997 Jun;13(3):179-89. doi: 10.1023/a:1005871614387. Int J Card Imaging. 1997. PMID: 9220280 Review.
-
The intravascular contrast agent Clariscan (NC 100150 injection) for 3D MR coronary angiography in patients with coronary artery disease.MAGMA. 2000 Nov;11(1-2):65-7. doi: 10.1007/BF02678498. MAGMA. 2000. PMID: 11186991 Clinical Trial. No abstract available.
-
Spatial factors for quantifying constant flow velocity in a small tube phantom: comparison of phase-contrast cine-magnetic resonance imaging and the intraluminal Doppler guidewire method.Jpn J Radiol. 2009 Nov;27(9):335-41. doi: 10.1007/s11604-009-0349-9. Epub 2009 Nov 27. Jpn J Radiol. 2009. PMID: 19943143
-
Magnetic resonance coronary angiography using navigator echo gated real-time slice following.Int J Card Imaging. 1998 Jun;14(3):191-9. doi: 10.1023/a:1006088727033. Int J Card Imaging. 1998. PMID: 9813756
-
Two-dimensional coronary MRA: limitations and artifacts.Eur Radiol. 1996;6(3):312-25. doi: 10.1007/BF00180601. Eur Radiol. 1996. PMID: 8798000
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous