Measurement of respiratory acoustical signals. Comparison of sensors

Affiliations
Comparative Study

Measurement of respiratory acoustical signals. Comparison of sensors

H Pasterkamp et al. Chest. 1993 Nov.

Abstract

We assessed the performance of three air-coupled and four contact sensors under standardized conditions of lung sound recording. Recordings were obtained from three of the investigators at the best site on the posterior lower chest as determined by auscultation. Lung sounds were band-pass filtered between 100 and 2,000 Hz and sampled simultaneously with calibrated airflow at a rate of 10 kHz. Fourier techniques were used for power spectral analysis. Average spectra for inspiratory sounds at flows of 2 +/- 0.5 L/s were referenced against background noise at zero flow. Air-coupled and contact sensors had comparable maximum signal-to-noise ratios and gave similar values for most spectral parameters. Unexpectedly, less sensitivity (lower signal-to-noise ratio) at high frequencies was observed in the air-coupled devices. Sensor performance needs to be characterized in studies of lung sounds. We suggest that lung sound spectra should be averaged at known airflows over several breaths and that all measurements should be reported relative to sounds recorded at zero flow.

PubMed Disclaimer

Comment in

Publication types

LinkOut - more resources