Characteristics of frequency-of-seeing curves in normal subjects, patients with suspected glaucoma, and patients with glaucoma
- PMID: 8258511
Characteristics of frequency-of-seeing curves in normal subjects, patients with suspected glaucoma, and patients with glaucoma
Abstract
Purpose: The authors performed this study to determine factors that affect the characteristics of frequency-of-seeing curves in normal subjects, patients with suspected glaucoma, and patients with glaucoma.
Methods: The sample consisted of 70 subjects (22 normal subjects, 12 patients with suspected glaucoma, and 36 patients with glaucoma). A program was written to interface with the Humphrey Field Analyzer (Humphrey Instruments, San Leandro, CA) to measure frequency-of-seeing curves. The authors presented stimuli 8 dB either side of the estimated threshold in 1-dB intervals with five repetitions at each stimulus intensity. The authors tested four to six locations in each subject, with randomization of the stimulus intensity and location. Fixation was monitored with the Heijl-Krakau method. Using a probit program, the authors calculated the threshold and slope (estimated by the interquartile range) of each curve.
Results: The authors obtained 124 curves from the normal subjects, 71 from the patients with suspected glaucoma, and 183 from the patients with glaucoma. In all three groups, the slope of the frequency-of-seeing curve was correlated highly with the threshold or threshold deviation, although the correlation was significantly higher in the normal subjects compared with the patients with suspected glaucoma and patients with glaucoma, even after controlling for the range of the threshold and threshold deviation. For this reason, the authors found considerably different frequency-of-seeing curves, between subject groups and also within the group of patients with glaucoma, in locations with the same threshold.
Conclusions: There may be fundamental differences in areas of normal subjects and patients with glaucoma with similar thresholds or threshold deviations. These differences also may exist within patients with glaucoma.
Similar articles
-
The psychometric function and reaction times of automated perimetry in normal and abnormal areas of the visual field in patients with glaucoma.Invest Ophthalmol Vis Sci. 1996 Apr;37(5):878-85. Invest Ophthalmol Vis Sci. 1996. PMID: 8603872
-
Variability in patients with glaucomatous visual field damage is reduced using size V stimuli.Invest Ophthalmol Vis Sci. 1997 Feb;38(2):426-35. Invest Ophthalmol Vis Sci. 1997. PMID: 9040476
-
Test-retest variability of frequency-doubling perimetry and conventional perimetry in glaucoma patients and normal subjects.Invest Ophthalmol Vis Sci. 1999 Mar;40(3):648-56. Invest Ophthalmol Vis Sci. 1999. PMID: 10067968
-
Threshold and variability properties of matrix frequency-doubling technology and standard automated perimetry in glaucoma.Invest Ophthalmol Vis Sci. 2005 Jul;46(7):2451-7. doi: 10.1167/iovs.05-0135. Invest Ophthalmol Vis Sci. 2005. PMID: 15980235
-
The properties of perimetric thresholds in normal and glaucomatous eyes.Ger J Ophthalmol. 1992;1(2):79-85. Ger J Ophthalmol. 1992. PMID: 1477630
Cited by
-
Detection and measurement of clinically meaningful visual field progression in clinical trials for glaucoma.Prog Retin Eye Res. 2017 Jan;56:107-147. doi: 10.1016/j.preteyeres.2016.10.001. Epub 2016 Oct 20. Prog Retin Eye Res. 2017. PMID: 27773767 Free PMC article. Review.
-
Advanced Vision Analyzer-Virtual Reality Perimeter: Device Validation, Functional Correlation and Comparison with Humphrey Field Analyzer.Ophthalmol Sci. 2021 Jun 25;1(2):100035. doi: 10.1016/j.xops.2021.100035. eCollection 2021 Jun. Ophthalmol Sci. 2021. PMID: 36249304 Free PMC article.
-
A novel Bayesian adaptive method for mapping the visual field.J Vis. 2019 Dec 2;19(14):16. doi: 10.1167/19.14.16. J Vis. 2019. PMID: 31845976 Free PMC article.
-
Towards Patient-Tailored Perimetry: Automated Perimetry Can Be Improved by Seeding Procedures With Patient-Specific Structural Information.Transl Vis Sci Technol. 2013 May;2(4):3. doi: 10.1167/tvst.2.4.3. Epub 2013 May 31. Transl Vis Sci Technol. 2013. PMID: 24049720 Free PMC article.
-
Frequency-of-seeing curves (psychometric functions) for perimetric stimuli in age-related macular degeneration.Ophthalmic Physiol Opt. 2025 Jan;45(1):301-307. doi: 10.1111/opo.13396. Epub 2024 Sep 27. Ophthalmic Physiol Opt. 2025. PMID: 39329222 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous