The Australian Incident Monitoring Study. Problems with ventilation: an analysis of 2000 incident reports
- PMID: 8273885
- DOI: 10.1177/0310057X9302100521
The Australian Incident Monitoring Study. Problems with ventilation: an analysis of 2000 incident reports
Abstract
A review of the first 2000 incidents reported to the Australian Incident Monitoring Study found 317 incidents which involved problems with ventilation. The major portion (47%) were disconnections; 61% of these were detected by a monitor. Monitor detection was by a low circuit pressure alarm in 37% but this alarm failed to warn of non-ventilation in 12 incidents (in 6 because it was not switched "on" and in 6 because of a failure to detect the disconnection). Failure of detection was usually with ventilator bellows descending in expiration. Complete failure to ventilate occurred in 143 incidents, most commonly because of a disconnection. Disconnection was associated, in one-third of the cases, with interference to the anaesthetic circuit by a third party and in nearly half with surgery on the head and neck. Leaks affected ventilation in 129 incidents, but in only 19 was ventilation totally lost; leaks associated with seal failure of the absorber were common. Misconnections occurred in 36 incidents, most commonly involving the scavenging system. The frequency of a complete failure to check an anaesthetic machine was greater when an induction room was involved than when only the operating theatre was the site of the incident. These incidents suggest that meticulous checking and monitoring for failure of ventilation, preferably using at least two separate, self-activating systems is highly desirable. The Australian and New Zealand College of Anaesthetists' policy on low circuit pressure alarms, oximetry and capnography is vindicated by these reports.
Similar articles
-
The Australian Incident Monitoring Study. Which monitor? An analysis of 2000 incident reports.Anaesth Intensive Care. 1993 Oct;21(5):529-42. doi: 10.1177/0310057X9302100508. Anaesth Intensive Care. 1993. PMID: 8273872
-
The Australian Incident Monitoring Study. The capnograph: applications and limitations--an analysis of 2000 incident reports.Anaesth Intensive Care. 1993 Oct;21(5):551-7. doi: 10.1177/0310057X9302100510. Anaesth Intensive Care. 1993. PMID: 8273874
-
The Australian Incident Monitoring Study. The oxygen analyser: applications and limitations--an analysis of 200 incident reports.Anaesth Intensive Care. 1993 Oct;21(5):570-4. doi: 10.1177/0310057X9302100513. Anaesth Intensive Care. 1993. PMID: 8273877
-
[Disconnections in the breathing system].Anaesthesist. 1993 Dec;42(12):835-46. Anaesthesist. 1993. PMID: 8304579 Review. German.
-
[Seal failure of the respiratory circuit during artificial ventilation of the lungs].Anesteziol Reanimatol. 1986 Mar-Apr;(2):69-74. Anesteziol Reanimatol. 1986. PMID: 3521398 Review. Russian. No abstract available.
Cited by
-
Composition and risk assessment of perioperative patient safety incidents reported by anesthesiologists from 2009 to 2019: a single-center retrospective cohort study.BMC Anesthesiol. 2021 Jan 7;21(1):8. doi: 10.1186/s12871-020-01226-0. BMC Anesthesiol. 2021. PMID: 33413123 Free PMC article.
-
Critical incidents during anesthesia in a developing country: A retrospective audit.Anesth Essays Res. 2010 Jul-Dec;4(2):64-8. doi: 10.4103/0259-1162.73508. Anesth Essays Res. 2010. PMID: 25885231 Free PMC article.
-
Enhanced notification of critical ventilator events.J Am Med Inform Assoc. 2005 Nov-Dec;12(6):589-95. doi: 10.1197/jamia.M1863. Epub 2005 Jul 27. J Am Med Inform Assoc. 2005. PMID: 16049226 Free PMC article.
-
Critical incident reporting in anaesthesia: a prospective internal audit.Indian J Anaesth. 2009 Aug;53(4):425-33. Indian J Anaesth. 2009. PMID: 20640204 Free PMC article.
-
New technology in anaesthesia: friend or foe?J Clin Monit Comput. 2014 Apr;28(2):113-6. doi: 10.1007/s10877-014-9557-7. J Clin Monit Comput. 2014. PMID: 24458593 No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources