Comparative Study
. 1993 Nov;2(6):474-7.

A stabilization period of 5 minutes is adequate when measuring pulmonary artery pressures after turning

Affiliations
  • PMID: 8275153
Comparative Study

A stabilization period of 5 minutes is adequate when measuring pulmonary artery pressures after turning

P A Shinners et al. Am J Crit Care. 1993 Nov.

Abstract

Objective: To compare hemodynamic measurements made before turning and at 5 and 30 minutes after turning, and to determine whether the stabilization period affects the difference between supine and side-lying pulmonary artery pressures.

Methods: This study was performed in the cardiothoracic surgical intensive care unit of a midwestern university hospital. The 31 postoperative open-heart surgical patients, 26 men and 5 women aged 41 to 76 years (64 +/- 9.3, mean +/- SD) with pulmonary artery catheters in place, were divided into two groups to compare supine to side-lying pressures and the time intervals between the position changes. The supine-first subjects (Group A) were placed in the supine position for baseline measurements and turned to either the right or left side-lying position for the 5- and 30-minute pulmonary artery pressure measurements. The side-first subjects (Group B) were placed in either the right or left side-lying position for baseline measurements and then in the supine position for the 5- and 30-minute pulmonary artery pressure measurements.

Results: Pulmonary artery pressures, heart rate and arterial pressure were not significantly different at 5 and 30 minutes. Supine pulmonary artery pressures in Group A were not significantly different from supine pressures in Group B. Side-lying pulmonary artery pressures in Group A were not significantly different from side-lying pressures in Group B. Side-lying vs supine pulmonary artery pressures were significantly different in both Group A and Group B.

Conclusion: The current practice of turning and settling the patient, zeroing the transducer and proceeding to make the pulmonary artery pressure readings appears to be valid. The stabilization period after turning does not explain the differences found between side-lying and supine pulmonary artery pressures.