Prevention of dislodgement of Cotrel-Dubousset rods from tulip screws
- PMID: 8278855
- DOI: 10.1097/00007632-199311000-00032
Prevention of dislodgement of Cotrel-Dubousset rods from tulip screws
Abstract
The authors report a case of posterior fusion of the lumbar spine using Cotrel-Dubousset instrumentation in which the rods dislodged from the tulip-headed pedicle screws postoperatively. They recommend that, to avoid this complication, particular care be taken to ensure the locking washers are inserted tightly without cross threading and that an adequate length of rod is left protruding beyond the screw.
Similar articles
-
Cotrel-Dubousset pedicle screw system for various spinal disorders. Merits and problems.Spine (Phila Pa 1976). 1991 Nov;16(11):1298-304. doi: 10.1097/00007632-199111000-00010. Spine (Phila Pa 1976). 1991. PMID: 1750004
-
Mid- to long-term outcomes in adolescent idiopathic scoliosis after instrumented posterior spinal fusion: a meta-analysis.Spine (Phila Pa 1976). 2013 Jan 15;38(2):E113-9. doi: 10.1097/BRS.0b013e31827ae3d0. Spine (Phila Pa 1976). 2013. PMID: 23124268 Review.
-
1997 Volvo Award winner in clinical studies. The effect of pedicle screw instrumentation on functional outcome and fusion rates in posterolateral lumbar spinal fusion: a prospective, randomized clinical study.Spine (Phila Pa 1976). 1997 Dec 15;22(24):2813-22. doi: 10.1097/00007632-199712150-00004. Spine (Phila Pa 1976). 1997. PMID: 9431617 Clinical Trial.
-
Comparison between single-screw and triangulated, double-screw fixation in anterior spine surgery. A biomechanical test.Spine (Phila Pa 1976). 1996 Dec 1;21(23):2728-34. doi: 10.1097/00007632-199612010-00005. Spine (Phila Pa 1976). 1996. PMID: 8979318
-
Grade I lumbar degenerative spondylolisthesis: do we always need fusion with bilateral screws and rods?J Neurosurg Sci. 2014 Jun;58(2 Suppl 1):65-71. J Neurosurg Sci. 2014. PMID: 25371951 Review. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical